This is the comprehensive '''User Manual''' for the VSParticle Generator One Spark Generator (VSP-G1). The [[G1_Quick_Start_Guide|Quick Start Guide]] is available as a quick reference for standard operation of the VSP-G1 unit.  
This is the comprehensive '''User Manual''' for the VSParticle Generator One Spark Generator (VSP-G1). The [[G1_Quick_Start_Guide|Quick Start Guide]] is available as a quick reference for standard operation of the VSP-G1 unit.  
<span style="background:#FF00FF">After completing the User Manual, cross-check the residual risks requirements...if any changes are made to information in the residual risks table (8.5), make sure the corresponding changes are made in the design document.</span>


{|class="table table-striped table-bordered"
{|class="table table-striped table-bordered"
Line 10: Line 7:
|-
|-
!Address  
!Address  
| Molengraaffsingel 10<br/>2629 JD<br/>Delft<br/>The Netherlands
| Oostsingel 209<br/>2612 HL<br/>Delft<br/>The Netherlands
|-
|-
!Website  
!Website  
Line 23: Line 20:
{{Warning||Before operating the VSP-G1 unit, please read&mdash;and understand&mdash;this manual.}}
{{Warning||Before operating the VSP-G1 unit, please read&mdash;and understand&mdash;this manual.}}


Wherever a caution [[File:Achtung.png|18px|frameless|CAUTION]] symbol is affixed to the machinery, there are explanatory texts located within the manual (see [[VSP-G1 Detailed System Description#Guide to symbols found on the VSP-G1|this section of the detailed system description for a guide to all symbols found on the VSP-G1.]] ). Other important safety information located within the manual are also labelled with the caution [[File:Achtung.png|18px|frameless|CAUTION]] symbol.
Wherever a caution [[File:Achtung.png|18px|frameless|CAUTION]] symbol is affixed to the machinery, there are explanatory texts located within the manual (see [[VSP-G1 Detailed System Description#Guide to symbols found on the VSP-G1 |this section of the detailed system description for a guide to all symbols found on the VSP-G1.]] ). Other important safety information located within the manual are also labelled with the caution [[File:Achtung.png|18px|frameless|CAUTION]] symbol.


You will also find conformity and warranty information at the very end of this user manual.
You will also find conformity and warranty information at the very end of this user manual.
Line 59: Line 56:
==== Operator safety actions ====
==== Operator safety actions ====
You are required to ensure that:<br>
You are required to ensure that:<br>
* the reactor is correctly mounted (see [[VSP-G1_User_Manual#Flow configurations|Flow configurations]]);  
* the reactor is correctly mounted (see [[VSP-G1_User_Manual#Setting_up|setting up]]);  
* there are no objects near the electrodes (see [[VSP-G1_User_Manual#Flow configurations|Flow configurations]]);
* there are no objects near the electrodes;
* process connections are correctly made (see [[VSP_G1_Process_Connections|Process connections]]);
* process connections are correctly made (see [[VSP-G1_User_Manual#Process_connections|process connections]]);
* the pressure in the system is limited to 1&plusmn;0.2&nbsp;bar absolute (see  [[VSP_G1_Process_Connections|Process connections]]);
* the pressure in the system is limited to 1&plusmn;0.2&nbsp;bar absolute (see  [[VSP-G1_User_Manual#Gas_pressure_monitoring |gas pressure monitoring]]);
* all, if any, external safety systems are functional (where applicable); and
* all, if any, external safety systems are functional (where applicable); and
* personal protective equipment are being used correctly (gloves, glasses, and FFP3/P3 certified filter mask).
* personal protective equipment are being used correctly (gloves, glasses, and FFP3/P3 certified filter mask).
Line 77: Line 74:
'''2. Reactor interlock'''
'''2. Reactor interlock'''
[[File:Reactor_interlock.png|thumb|The mounting pin is used both to secure the reactor assembly, and to close the interlock protection.]]
[[File:Reactor_interlock.png|thumb|The mounting pin is used both to secure the reactor assembly, and to close the interlock protection.]]
The VSP-G1 detects whether the reactor endcaps are properly mounted, using two mounting pins (see image). You must ensure that the reactor itself is correctly assembled. In short, this means that the reactor is leak tight (see [[VSP-G1_User_Manual#Leak tightness protocol|Leak tightness protocol]]) and that the gas inlet valve and electrode tips are free to move with a minimum clearance of 10&nbsp;mm from any foreign objects (see [[VSP-G1_User_Manual#Flow configurations|Flow configurations]]).
The VSP-G1 detects whether the reactor endcaps are properly mounted, using two mounting pins (see image). You must ensure that the reactor itself is correctly assembled. In short, this means that the reactor is leak tight (see [[VSP-G1_Operating_and_maintenance_guide|the operating and maintenance guide]]) and that the gas inlet valve and electrode tips are free to move with a minimum clearance of 10&nbsp;mm from any foreign objects (see [[VSP-G1_User_Manual#Process connections|the section on process connections]] and [[VSP-G1_User_Manual#Flow_configurations|the section on flow configurations]]).


{{Warning||Gas connections and other user-provided functions are not controlled by the interlock. An optional external interlock is available for the VSP-G1.}}
{{Warning||Gas connections and other user-provided functions are not controlled by the interlock. An optional external interlock is available for the VSP-G1.}}
Line 89: Line 86:
''Non-ionizing radiation.'' The spark plasma emits a bright light in the visible and UV range. A viewport with a removable optical filter is included to allow visual inspection of the electrodes. Additionally, electromagnetic radiation (radio frequency) may be emitted.
''Non-ionizing radiation.'' The spark plasma emits a bright light in the visible and UV range. A viewport with a removable optical filter is included to allow visual inspection of the electrodes. Additionally, electromagnetic radiation (radio frequency) may be emitted.


{{Notification||See [[VSP-G1_User_Manual#Residual_risks|Residual risks]] for a comprehensive list of residual risks and how they are mitigated through design features and/or documentation.}}
<!-- {{Notification||See [[VSP-G1_User_Manual#Residual_risks|Residual risks]] for a comprehensive list of residual risks and how they are mitigated through design features and/or documentation.}} -->


=== Safety of nanoparticles ===
=== Safety of nanoparticles ===
Because our spark generators can in principle be used to produce nanoparticles from any metal, we start from the assumption that these nanoparticles pose potential risks to the health of operators, consumers and to the environment. This manual addresses operator exposure. Consumer exposure is determined by the application in which the nanomaterials are to be used and should be evaluated on a case-by-case basis by the user. Environmental considerations should also be addressed by the operators with a [[VSP-G1_User_Manual#Life cycle analysis|life cycle analysis]].
Because our spark generators can in principle be used to produce nanoparticles from any metal, we start from the assumption that these nanoparticles pose potential risks to the health of operators, consumers and to the environment. This manual addresses operator exposure. Consumer exposure is determined by the application in which the nanomaterials are to be used and should be evaluated on a case-by-case basis by the user. Environmental considerations should also be addressed by the operators with a [[VSP-G1_Background_information#Lifecycle_analysis|life cycle analysis]].


The main channels of operator exposure to nanoparticles are through (skin) absorption, ingestion and inhalation. The aerosol nature of the produced nanoparticles by the VSP-G1 system means exposure through inhalation is the main risk. The respiratory system of the human body has natural defenses in place for airborne particles such as dust, viruses and bacteria. However, nanoparticles are able to bypass most of these defenses due to their small size. For this reason, nanoparticles are treated in the same way as fine particles. More information on health and safety of nanoparticles can be found through the [[VSP-G1_User_Manual#Links to more information on nanoparticles and health|links at the end of this section]].
The main channels of operator exposure to nanoparticles are through (skin) absorption, ingestion and inhalation. The aerosol nature of the produced nanoparticles by the VSP-G1 system means exposure through inhalation is the main risk. The respiratory system of the human body has natural defenses in place for airborne particles such as dust, viruses and bacteria. However, nanoparticles are able to bypass most of these defenses due to their small size. For this reason, nanoparticles are treated in the same way as fine particles. More information on health and safety of nanoparticles can be found through the [[VSP-G1_User_Manual#Links to more information on nanoparticles and health|links at the end of this section]].
Line 103: Line 100:
! Exposure risk situation !! Operator precautions
! Exposure risk situation !! Operator precautions
|-
|-
| Gas phase exposure || An aerosol system must be leak tight from source to deposition (see [[VSP-G1_User_Manual#Leak tightness protocol|Leak tightness protocol]]).
| Gas phase exposure || An aerosol system must be leak tight from source to deposition (see section 2.1 of the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]]).
|-
|-
| Filtration || The gas exhaust should be treated with a suitable HEPA or ULPA filter. VSParticle recommends using a large capacity HEPA pre-filter, followed by an ULPA end filter.
| Filtration || The gas exhaust should be treated with a suitable HEPA or ULPA filter. VSParticle recommends using a large capacity HEPA pre-filter, followed by an ULPA end filter.
|-
|-
| Dealing with pyrophoric materials || Metal nanoparticles can be extremely reactive, and can be pyrophoric (spontaneous ignition upon exposure to air).Controlled air-exposure, however, is needed before opening the system in order to passivate any reactive materials (see [[VSP-G1_User_Manual#Controlled-air exposure protocol|Controlled-air exposure protocol]]).
| Dealing with pyrophoric materials || Metal nanoparticles can be extremely reactive, and can be pyrophoric (spontaneous ignition upon exposure to air).Controlled air-exposure, however, is needed before opening the system in order to passivate any reactive materials (see [[VSP-G1_Operating_and_maintenance_guide#Controlled_air_exposure|controlled-air exposure protocol]]).
|-
|-
| Product handling || When the system must be opened to insert or remove a substrate, methods as described in the [[VSP-G1 Maintenance protocols | maintenance protocols section]] must be followed to reduce the risk of nanoparticle exposure.
| Product handling || When the system must be opened to insert or remove a substrate, methods as described in the [[VSP-G1 Maintenance protocols | maintenance protocols section]] must be followed to reduce the risk of nanoparticle exposure.
|-
|-
| Maintenance || Nanoparticles will collect on the surfaces inside the reactor and downstream piping. Cleaning of these surfaces presents the largest exposure risk to operators, unless they can be performed in a safe environment (e.g. a glovebox) (see [[VSP-G1_User_Manual#System care|System care]] for the relevant maintenance protocols).  
| Maintenance || Nanoparticles will collect on the surfaces inside the reactor and downstream piping. Cleaning of these surfaces presents the largest exposure risk to operators, unless they can be performed in a safe environment (e.g. a glovebox) (see [[VSP-G1_User_Manual#System care|system care]] and the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]] for the relevant maintenance protocols).  
<br />
<br />
There is also a separate protocol for dismantling and sealing used filters for disposal/recycling, which is largely dependent on your particular filtration system (see [[VSP-G1 Maintenance protocols#Filter replacement|Filter replacement in the maintenance protocols section]]).
There is also a separate protocol for dismantling and sealing used filters for disposal/recycling, which is largely dependent on your particular filtration system (see the section on filter replacement in the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]]).
|}
|}


Line 120: Line 117:
* [http://www.veiligheidskunde.nl/xu/document/cms/streambin.asp?requestid=C21915B8-EB30-4FC1-B2B3-14CA09C2E9FE Nanosafety Guidelines], a set of safety recommendations by the workgroup Nanosafety of the Faculty of Applied Sciences of Delft University of Technology for working with ‘free nanostructured matter’ in research activities.
* [http://www.veiligheidskunde.nl/xu/document/cms/streambin.asp?requestid=C21915B8-EB30-4FC1-B2B3-14CA09C2E9FE Nanosafety Guidelines], a set of safety recommendations by the workgroup Nanosafety of the Faculty of Applied Sciences of Delft University of Technology for working with ‘free nanostructured matter’ in research activities.
* [https://www.sofokles.nl/wp-content/uploads/Part_B-Good_Practices_Use_gineered_Nanomaterials_Dutch_Academic.pdf Good practices for the use of engineered nanomaterials in a research environment], a guide describing good practices for working with engineered nanomaterials by FOM/NWO (The Netherlands Organisation for Scientific Research).
* [https://www.sofokles.nl/wp-content/uploads/Part_B-Good_Practices_Use_gineered_Nanomaterials_Dutch_Academic.pdf Good practices for the use of engineered nanomaterials in a research environment], a guide describing good practices for working with engineered nanomaterials by FOM/NWO (The Netherlands Organisation for Scientific Research).
* Nanosafety Quick Check, a checklist developed based on a summary of the Nanosafety guidelines listed above <span style="background:#00FF00"> is there a link to this, or can we provide this as a downloadable file? permissions? </span>
* [https://www.veiligheidskunde.nl/xu/document/cms/streambin.asp?requestid=C21915B8-EB30-4FC1-B2B3-14CA09C2E9FE Nanosafety Quick Check], a checklist developed based on a summary of the Nanosafety guidelines listed above.
* [http://www.nanowerk.com/nanotechnology/ten_things_you_should_know_9.php Ten things you should know about nanotechnology], an article available on Nanowerk website
* [http://www.nanowerk.com/nanotechnology/ten_things_you_should_know_9.php Ten things you should know about nanotechnology], an article available on Nanowerk website


Line 141: Line 138:


=== Detailed system description ===
=== Detailed system description ===
A visual reference and detailed description of the components that are part of the VSP-G1 system is provided on the [[VSP-G1 Detailed System Description | Detailed System Description]] page. This includes a [[VSP-G1 Detailed System Description#Guide to symbols found on the VSP-G1|full reference of the warning labels and symbols]] found on the machine.
A visual reference and detailed description of the components that are part of the VSP-G1 system is provided on the [[VSP-G1 Detailed System Description | Detailed System Description]] page. This includes a [[VSP-G1 Detailed System Description#Symbols|full reference of the warning labels and symbols]] found on the machine.


== Setting up ==
== Setting up ==
Line 156: Line 153:
}}
}}


{{Warning|Do not open the base unit. | There are no user serviceable parts inside.}}
{{Warning|Do not open the base unit or the endcaps. | There are no user serviceable parts inside.}}


{|class="table"
{|class="table"
Line 166: Line 163:
=== Mounting the reactor assembly ===
=== Mounting the reactor assembly ===
{{Warning||Under no circumstances mount the endcaps onto the base unit without a sealed reactor.}}
{{Warning||Under no circumstances mount the endcaps onto the base unit without a sealed reactor.}}
When shipped, the reactor assembly is attached to the transport plate in the packaging. The electrodes have already been installed. In the case you want to replace the electrodes, please see the electrode replacement protocol in the [[VSP-G1_User_Manual#System Care|System care]] section of this manual.
When shipped, the reactor assembly is attached to the transport plate in the packaging. The electrodes have already been installed. In the case you want to replace the electrodes, please see the electrode replacement protocol in the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]].


{| class="table table-bordered table-striped"
{| class="table table-bordered table-striped"
Line 172: Line 169:
! Step !! Description !! Picture
! Step !! Description !! Picture
|-
|-
| 1. || Unscrew and remove the alignment pins holding the reactor assembly to the transport plate. || [[Image:Alignment pins.png|thumb]]
| 1. || Place the reactor assembly and transport plate close to the VSP-G1 base unit and unscrew and remove the alignment pins holding the reactor assembly to the transport plate. || [[Image:Alignment pins.png|thumb]]
|-
|-
| 2.|| Holding the endcaps in both hands, carefully transfer the reactor assembly to the mount plate located at the top of the base unit, aligning the markings on the bottom of the reactor assembly's endcaps and the mount plate. || [[Image:Mounting the reactor assembly.png|thumb]]
| 2.|| Holding the endcaps in both hands, carefully transfer the reactor assembly to the mount plate located at the top of the base unit, aligning the markings on the bottom of the reactor assembly's endcaps and the mount plate. || [[Image:Mounting the reactor assembly.png|thumb]]
Line 188: Line 185:
Insert the alignment pins and tighten.
Insert the alignment pins and tighten.
|}
|}
[[Image:Ground check.JPG|thumb|Checking the resistance between the reactor chamber and the ground point.]]
To confirm that the reactor assembly is correctly assembled, use a digital multimeter (DMM) to measure the resistance between the reactor chamber and the ground point on the back of the system. The resistance should be <0.1 Ohm (keep in mind that the leads in a typical handheld DMM can have a resistance as high as 0.5 Ohm).


To confirm that the reactor assembly is correctly assembled, use a digital multimeter (DMM) to measure the resistance between the reactor chamber and the ground point on the back of the system. <mark>Image needed.</mark> The resistance should be <0.1 Ohm.
=== Building a basic setup ===  
 
=== Building a simple setup ===
=== Your customised setup ===  
Your particular deposition requirements and research needs will result in a system setup that is different from others. In all cases, however, the following requirements help ensure that the system works efficiently and safely:
Your particular deposition requirements and research needs will result in a system setup that is different from others. In all cases, however, the following requirements help ensure that the system works efficiently and safely:
* Working system that allows compartmentalisation of the reactor and the deposition system;
* Use correctly rated tools and components
* Gas pressure monitoring tool;
* Apply compartmentalisation of the reactor and the deposition system;
* Deposition system; and
* Use a gas pressure monitoring tool;
* Safe exhaust (fume hood with HEPA+ filters, for example) for the outflowing gas.
* Use a safe exhaust (fume hood with HEPA+ filters, for example) for the outflowing gas; and
* Perform regular leak checks.


 
<!-- A guide to a reference setup is available to help designing basic setup that fulfills these requirements on the [[VSP-G1 Reference setup]] page.
A guide to a reference setup is available to help designing basic setup that fulfills these requirements on the [[VSP-G1 Reference setup]] page.
<mark>Move to Reference setup page</mark> -->
<mark>Move to Reference setup page</mark>


[[File:Simple set up.png|thumb|right|''A simple setup connecting the input gas source and flow system to the particle source, deposition system, filtration system and the ventilation system.'']]
[[File:Simple set up.png|thumb|right|''A simple setup connecting the input gas source and flow system to the particle source, deposition system, filtration system and the ventilation system.'']]
The following basic operating procedure is for a simple setup pictured below, which comprises several operating units: a gas source and flow system, a particle source (the VSP-G1 unit), a pressure gauge, a deposition system and exhaust filters. To complete the setup, the units are connected with tubes, piping, fittings, valves and so on. The ball valves ([[File:ball valve.png|20px]]) in this setup help compartmentalize the overall system for leak testing and for safe disconnection. Each setup will be different, depending on your individual needs, but the concepts described here should generally be valid.
The following basic operating procedure is for the simple setup pictured below, which comprises several operating units: a gas source and flow system; a particle source (the VSP-G1 unit); a pressure gauge; a deposition system; and exhaust filters. To complete the setup, the units are connected with tubes, piping, fittings, valves and so on. The ball valves ([[File:ball valve.png|20px]]) in this setup help compartmentalize the overall system for leak testing and for safe disconnection. Each setup will be different, depending on your individual needs, but the concepts described here should generally be valid.


To reduce particle losses, please take the following guidelines into account:
To reduce particle losses, please take the following guidelines into account:
* Keep the tubing length between particle source and deposition system as short as possible to reduce diffusive losses.
* Keep the tubing length between particle source and deposition system as short as possible to reduce diffusive losses.
* Use conductive tubing (metal or conductive polymer) in particle carrying lines to reduce electrostatic losses.
* Use conductive tubing (metal or conductive polymer (electrostatic dissipative tubing) in particle carrying lines to reduce electrostatic losses.
* Do not use flow controllers (e.g. needle valves, mass flow controller) in particle lines, as particles will clog up the controller.
* Do not use flow controllers (e.g. needle valves, mass flow controller) in particle lines, as particles will clog up the controller.
** Usually the controller can be placed in another (particle free) line, or downstream of the deposition system, behind a filter.
** Usually the controller can be placed in another (particle free) line, or downstream of the deposition system, behind a filter.
Line 215: Line 211:
** Use ball valves with a wide bore.
** Use ball valves with a wide bore.
** Avoid fibers and meshes.
** Avoid fibers and meshes.
<mark>TC: IS THERE A CODE FOR AUTOMATICALLY RESIZING THE IMAGE BASED ON BROWSER SIZE? USEFUL FOR OPERATORS ACCESSING MANUAL USING TABLETS IN THE LAB </mark>


Building a setup takes three steps:
Building a setup takes three steps:
Line 226: Line 218:
# Check the system for leak tightness and make adjustments if necessary. (See [[VSP-G1_User_Manual#Checking for leaks|Checking for leaks]].)
# Check the system for leak tightness and make adjustments if necessary. (See [[VSP-G1_User_Manual#Checking for leaks|Checking for leaks]].)


==== Compartmentalisation ====
==== Process connections ====
[[file:leak tightness compartmentalisation.png|thumb|right|''Compartmentalisation of the setup.'']]
Use only tubing and components rated for the operating conditions of the system. When using quick-connect fittings and other systems that do not require tools, ensure that the system is protected against accidental opening (e.g. clamps, latches and/or signs).
For safety reasons, we recommend compartmentalisation of every unit within the setup that regularly opens up to the atmosphere to allow for easy dismantling in a safe manner. In the setup described here, we use two valves on both sides of the deposition chamber which allows us to easily seal off both the reactor and the rest of the setup. When the deposition chamber is removed, it can be safely transported to a safe working location, such as a fumehood or glovebox, in order to take out the sample. Meanwhile, the rest of the setup, which often contains a significant amount of nanomaterial, remains safely sealed.
Different technologies can be used to connect tubing and the other components. The process connections must be made according to the fitting vendor's specifications.  The VSP-G1 comes equipped with [https://www.swagelok.com/downloads/webcatalogs/EN/MS-13-151.PDF Swagelok®] fittings.  


==== Flow configurations ====
{{Warning||Do not use a rigid gas connection on the gas inlet. Allow for 25mm movement of the reactor inlet tube. Damage may occur if the inlet tube is not free to move during the electrode alignment procedure.
<mark>content needed</mark>
{{Warning||Do not use a rigid gas connection on the gas inlet. Allow for 25mm movement of the reactor inlet tube. Damage may occur if the inlet tube is not free to move during the electrode homing sequence.
}}
{{Warning||Pressure may build up in the reactor if the hollow electrode is the only outlet, and one of the side inlets is used as an inlet. In such configurations, the user must ensure pressure cannot rise above 1.2 bar absolute.
}}
}}


==== Process connections ====
{{Warning|Important notes about carrier gases:|
{{Warning|Important notes about carrier gases:|
* Do not use reactive gases, in particular gases containing more than a few percent oxygen, such as air.
* Do not use reactive gases, in particular gases containing more than a few percent oxygen, such as air.
* Do not use heated gases.}}
* The VSP-G1 is designed for use at room temperature. Do not use heated gases.}}
 
<!-- <mark>Here, we'll discuss a simple set-up using one inlet and outlet using Swagelok connectors. On [[VSP G1 Process Connections| this page]] you'll find much more information on the available configurations and connections.</mark> -->


<mark>Here, we'll discuss a simple set-up using one inlet and outlet using Swagelok connectors. On [[VSP G1 Process Connections| this page]] you'll find much more information on the available configurations and connections.</mark>
==== Compartmentalisation ====
[[file:leak tightness compartmentalisation.png|thumb|right|''Compartmentalisation of the setup.'']]
For safety reasons, we recommend compartmentalisation of every unit within the setup that regularly opens up to the atmosphere to allow for easy dismantling in a safe manner. In the setup described here, we use two valves on both sides of the deposition chamber which allows us to easily seal off both the reactor and the rest of the setup. When the deposition chamber is removed, it can be safely transported to a safe working location, such as a fumehood or glovebox, in order to take out the sample. Meanwhile, the rest of the setup, which often contains a significant amount of nanomaterial, remains safely sealed.


==== Pressure safety ====
==== Gas pressure monitoring ====
{{Warning|| Do not exceed +1.2 bar absolute. Do not turn on the spark below 0.8 bar.}}
{{Warning|| Do not exceed 1.2 bar absolute. Do not turn on the spark below 0.8 bar.}}


The internal pressure should not exceed +1.2 bar or there is an increased risk of failure of the gaskets and/or seals in the reactor, causing leaks and risk of nanoparticle exposure. A build-up of pressure within the system might be an indicator of obstruction, either within the reactor, the deposition system or in the filtration system. Ideally, a pressure sensor at the reactor inlet is used to automatically switch off the gas supply. Stop spark production and gas flow immediately when pressure reaches +1.2 bar to determine the source of obstruction and resolve the issue. See [[VSP-G1_User_Manual#System Care|System care]] for cleaning procedures.
The internal pressure should not exceed +1.2 bar or there is an increased risk of failure of the gaskets and/or seals in the reactor, causing leaks and risk of nanoparticle exposure. A build-up of pressure within the system might be an indicator of obstruction, either within the reactor, the deposition system or in the filtration system. Ideally, a pressure sensor at the reactor inlet is used to automatically switch off the gas supply. Stop spark production and gas flow immediately when pressure reaches +1.2 bar to determine the source of obstruction and resolve the issue. See [[VSP-G1_User_Manual#System Care|System care]] for cleaning procedures.
==== Flow configurations ====
The VSP-G1 has multiple inlets and outlets, that allow the user to modify the flow profile around the spark gap. Configurations usually have a single outlet for the particles, but can have a single or multiple inlets. Possible reasons to modify the flow configuration are to change the dilution rate in the gap, or to apply a sheath or dilution gas. Particular attention must be taken if the hollow electrode is used as the only outlet: During the startup sequence, the VSP-G1 lets the electrodes touch, obstructing the outlet flow.
{{Warning||Pressure may build up in the reactor if the hollow electrode is the only outlet, and one of the side inlets is used as an inlet. In such configurations, the user must ensure pressure cannot rise above 1.2 bar absolute.
}}
{{Warning||When using the hollow electrode as the only outlet, be sure not to mount a solid electrode tip.
}}
==== Exhaust ====
The gas exhaust should be treated with a suitable HEPA or ULPA filter. VSParticle recommends using a large capacity HEPA pre-filter, followed by an ULPA end filter. Replace the pre-filter before the end-filter is filled. The filtered exhaust should not be vented directly into a workspace; vent the outflowing gas through a fume hood or other suitable (non-recirculating) ventilation exhaust.


==== Checking for leaks ====
==== Checking for leaks ====
{{Warning|| The leak tightness protocol is an essential part of the operating procedure before initiating particle production. Failure to test for leak tightness, improper installation/removal of blindcaps or lack of external monitoring tools could mean the total output of the system&mdash;that is, up to 100mg/h of nanoparticles&mdash;is released continuously to the operator.}}
{{Warning|| The leak tightness protocol is an essential part of the operating procedure before initiating particle production. Failure to test for leak tightness, improper installation/removal of blindcaps or lack of external monitoring tools could mean the total output of the system&mdash;that is, up to 100mg/h of nanoparticles&mdash;is released continuously to the operator.}}
Refer to the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]] for information on how to perform a leak test on the system.


=== Electronic connections ===
=== Electronic connections ===
Line 278: Line 282:
The VSP-G1 can be remotely controlled using RS232 over the sub-D connector. Please refer to the [[VSP-G1_User_Manual#Remote_controlled_operation | section 5.1.2 on remote controlled operation]].
The VSP-G1 can be remotely controlled using RS232 over the sub-D connector. Please refer to the [[VSP-G1_User_Manual#Remote_controlled_operation | section 5.1.2 on remote controlled operation]].


=== Final set up checklist ===
<!--=== Final set up checklist ===
<mark> Content needed (including ensuring that the system is grounded)</mark>
<mark> Content needed (including ensuring that the system is grounded)</mark>
<mark>move to operation?</mark>
<mark>move to operation?</mark>-->


== Operating the system ==
== Operating the system ==
Line 293: Line 297:
The VSP-G1 can be controlled directly from the front panel using the two dials and the button on the left hand side. The two dials on the front panel can be pressed to select a function, such as changing a set point. In general, the dials can be used to change the value shown directly above it on the display by pressing the dial and then rotating it to change the associated value. Another press of the dial then locks or confirms the value. The button on the left hand side of the front panel is used solely to toggle sparking. Refer to the [[VSP-G1_Detailed_System_Description#Controls_and_indicators|controls and indicators section of the detailed system description]] for additional information. This section further describes the key functions of the dials and button in various states of operation of the VSP-G1.
The VSP-G1 can be controlled directly from the front panel using the two dials and the button on the left hand side. The two dials on the front panel can be pressed to select a function, such as changing a set point. In general, the dials can be used to change the value shown directly above it on the display by pressing the dial and then rotating it to change the associated value. Another press of the dial then locks or confirms the value. The button on the left hand side of the front panel is used solely to toggle sparking. Refer to the [[VSP-G1_Detailed_System_Description#Controls_and_indicators|controls and indicators section of the detailed system description]] for additional information. This section further describes the key functions of the dials and button in various states of operation of the VSP-G1.


===== '''Starting the alignment sequence''' =====
===== Starting the alignment sequence =====
When powering up the device, the message "Begin electrode alignment" appears on the display. Confirm no carrier gas is flowing and that the gas inlet is free to move, then confirm by pressing the right hand dial.
When powering up the device, the message "Begin electrode alignment" appears on the display. Confirm no carrier gas is flowing and that the gas inlet is free to move, then confirm by pressing the right hand dial.
===== '''Changing the charging current and/or gap voltage set point''' =====
===== Changing the charging current and/or gap voltage set point =====
This can be done during sparking and when the "Ready to spark" message appears on the display. Press the dial underneath the value you wish to change and then rotate the dial to change the set point. Press the dial again or leave it untouched for 2 seconds to confirm the value.
This can be done during sparking and when the "Ready to spark" message appears on the display. Press the dial underneath the value you wish to change and then rotate the dial to change the set point. Press the dial again or leave it untouched for 2 seconds to confirm the value.
===== '''Start/stop sparking''' =====
===== Start/stop sparking =====
Sparking can only be initiated when the "Ready to spark" message is shown on the display. Press the button to start sparking. When sparking, another press of the button will return the system to the idle "Ready to spark" mode.
Sparking can only be initiated when the "Ready to spark" message is shown on the display. Press the button to start sparking. When sparking, another press of the button will return the system to the idle "Ready to spark" mode.
===== '''Accessing and navigating the menu''' =====
===== Accessing and navigating the menu =====
When the "Ready to spark" message is displayed, press and release both dials simultaneously to access the menu. By turning the right hand dial, you can browse the menu options. Press the right hand dial to confirm or alter a menu setting. When altering a setting, press the right hand dial again to confirm the value. Pressing the left hand dial exits the menu and returns the system to the "Ready to spark" mode.
When the "Ready to spark" message is displayed, press and release both dials simultaneously to access the menu. By turning the right hand dial, you can browse the menu options. Press the right hand dial to confirm or alter a menu setting. When altering a setting, press the right hand dial again to confirm the value. Pressing the left hand dial exits the menu and returns the system to the "Ready to spark" mode.


Line 305: Line 309:
{{Warning||Do not open the system while it is being operated remotely.
{{Warning||Do not open the system while it is being operated remotely.
}}
}}
The VSP-G1 can be controlled remotely via the 9 pin D-SUB connector at the back of the system. A full discussion of the command structure, functions and limitations can be found in the [[VSP-G1 Remote control guide|Remote control guide]].
The VSP-G1 can be controlled remotely via the 9 pin D-SUB connector at the back of the system. The remote control structure allows for changing set points and reading the status of the device. It does not allow access to the settings menu and cannot initiate the electrode alignment procedure during start up. A full discussion of the command structure, functions and limitations can be found in the [[VSP-G1 Remote control guide|Remote control guide]].


=== Basic operating procedure ===
=== Basic operating procedure ===
''An abridged form of this section is also available as a [https://docs.vsparticle.com/w/index.php/G1_Quick_Start_Guide'''Quick Start Guide.''']<br>
''An abridged form of this section is also available as a [https://docs.vsparticle.com/w/index.php/G1_Quick_Start_Guide'''Quick Start Guide'''], including a themed version for printing.
<mark>reference to other protocols?</mark>


====  Preparation ====
====  Preparation ====
# Ensure that the system is [[VSP-G1_User_Manual#Leak tightness protocol|leak tight]].
# Ensure that the system is [[VSP-G1_User_Manual#Checking_for_leaks|leak tight]].
# Power up all systems.
# Power up all systems, includeing the VSP-G1 using the mains switch on the back panel near the power connector. Ensure the power cable is properly fitted and powered.
# Ensure that the path to the exhaust is clear.
# When the "Start electrode alignment?" message appears on the display, press and release the right hand dial on the front panel to start the alignment sequence.
# Start gas flow.
# Wait for the alignment procedure to complete. Several messages appear on the display during the sequence to indicate activity. The integrated LED on the button will blink during the entire sequence and stays on once completed.
==== Start nanoparticle production ====
# Ensure that the path to the exhaust is clear and all required valves are open.
# Set target voltage and current using the dials on the front of the VSP-G1 base unit.
# Start the carrier gas flow.
# Press the start button to begin nanoparticle production.
# Monitor the pressure to ensure it is within the required range (0.8-1.2 bar(a)).
# To stop production, press the start button again.
 
# It's possible to adjust the particle output by changing the voltage and current settings. Higher voltage and current leads to larger particles.
{{Notification|Carrier gas selection:|The VSP-G1 supports Ar and N<sub>2</sub> as carrier gases. The system needs to be configured for the carrier gas that you use. Follow the instructions in [[VSP-G1_Operating_and_maintenance_guide#Changing_carrier_gas | the operating and maintenance guide]] to verify the right carrier gas setting is selected. If set incorrectly, the spark will be unstable or very slow in reaching the desired setpoints.}}
 
==== Start and stop nanoparticle production ====
The set points for charging current (frequency control) and gap voltage may be altered when the system displays the "Ready to spark" message and when the system is sparking. The set points are controlled by the two dials on the front panel. The left hand dial is associated with the gap voltage setpoint while the right hand dial controls the charging current. Higher voltage and current generally lead to larger particles.To change a setpoint:
 
#Press and release the corresponding dial. The unit of the associated value (e.g. kV or mA) will start blinking indicating the value is unlocked.
#Rotate the dial to set the desired value.
#Press and release the dial once more or wait 2 seconds to lock the value.
 
To start sparking, press and release the button on the left hand side of the front panel. The indicator LED around the button will change from solid to blinking red, indicating the machine is in operation. To stop sparking, press and release the button once more. This will return the system to the "Ready to spark" mode.
It's possible to adjust the particle output by changing the voltage and current settings.  


We recommend leaving the system to equilibrate/stabilise for 15-30 minutes, depending on your specific needs. This time allows the system to warm up, and the surface of the electrode tips to be refreshed and removing, if any, surface oxides.
We recommend leaving the system to equilibrate/stabilise for 15-30 minutes, depending on your specific needs. This time allows the system to warm up, and the surface of the electrode tips to be refreshed and removing, if any, surface oxides.
====  Shutting down ====
====  Shutting down ====
# Stop particle production.
# Make sure the system is in "Ready to spark" mode.
# Select Shutdown in the menu, and press 'OK'.
# Press and release both dials simultaneously to enter the menu. The option to shut down the system is selected upon reaching the menu.
# Turn off the mains switch.
# Press and release the right hand dial to confirm shutdown.
# Wait for the message "Please power off" to appear
# Set the mains switch near the power cable at the back of the system to the 0 position.


==== Opening the system ====
==== Opening the system ====
[[file:containing the deposition system for removal.png|thumb|''Four valves, two on each side of the deposition system have been closed to contain it for removal from the setup. When removing the deposition system, the contaminated surface that the operator can be exposed to is reduced to the volume between the valves.'']]Common reasons to open the system are to place or remove a sample/substrate, to clean the reactor, or to change electrodes. To reduce the risk of exposure when handling non-immobilized nanomaterials, it's recommended to open the system in a controlled working environment. If the deposition chamber is not equipped with sufficient protective measures, it can be convenient to move the deposition chamber itself to a fume-hood or glovebox (see image).
[[file:containing the deposition system for removal.png|thumb|''Four valves, two on each side of the deposition system have been closed to contain it for removal from the setup. When removing the deposition system, the contaminated surface that the operator can be exposed to is reduced to the volume between the valves.'']]Common reasons to open the system are to place or remove a sample/substrate, to clean the reactor, or to change electrodes. To reduce the risk of exposure when handling non-immobilized nanomaterials, it's recommended to open the system in a controlled working environment. If the deposition chamber is not equipped with sufficient protective measures, it can be convenient to move the deposition chamber itself to a fume-hood or glovebox (see image).


Steps required before opening the system:
Steps required before opening the system:
# Stop particle production.
# Stop particle production.
# Flush the system (see [[VSP-G1_User_Manual#Flushing the system|Flushing the system]]).
# Flush the system (see the corresponding section in the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]]).
# Stop the gas flow.
# Stop the gas flow.
# Take out your sample following the procedure for your deposition system, taking the necessary precautions to prevent/reduce exposure to nanoparticles. Pay special attention when handling potentially [[VSP-G1_User_Manual#Controlled air exposure|pyrophoric nanoparticles]].
# Take out your sample following the procedure for your deposition system, taking the necessary precautions to prevent/reduce exposure to nanoparticles. Pay special attention when handling potentially [[VSP-G1_Operating_and_maintenance_guide#Controlled air exposure|pyrophoric nanoparticles]].


== System care ==
== System care ==
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']


{{Warning|| Important points to remember when conducting maintenance activities involving disassembly and reassembly:
The system requires no regular maintenance by the end user, but cleaning the reactor periodically is recommended, especially after a change of electrode material. To prevent exposure to nanoparticles and to ensure safe and efficient cleaning, guidelines are set out in the [[VSP-G1_Operating_and_maintenance_guide| operating and maintenance guide]]. The [[VSP-G1_Operating_and_maintenance_guide#Replacing_electrodes|procedure to replace electrodes]] is also discussed there.
* Turn off the spark generator and disconnect it from the power supply;
* Transport the reactor assembly with the transport plate and open it in a safe environment;
* Do not apply pressure when wiping the surface;
* The coating on the viewport filter screens electromagnetic radiation. Do not scratch the coating, nor replace it with a different type;
* Dry all surfaces before reassembly;
* Retract the electrodes sufficiently to avoid collision with each other; and
* Screw down '''both''' reactor mounting pins.}}


===When to clean the reactor and replace the filter(s)===  
===When to clean the reactor and replace the filter(s)===  
Clean the reactor whenever you replace the electrodes, the reactor is dirty, or after 100-1000 hours of operation (which in turn depends on your operating conditions: if making more particles, clean more often). If higher settings are needed to maintain a spark flow, it is possible that the reactor needs to be cleaned to prevent spark obstruction which can possibly lead to spark misfiring within the reactor system. If the pressure within the system increases even with the valves open, it is possible that the filter is full and needs to be replaced.
Clean the reactor whenever you replace the electrodes, the reactor is dirty, or after 100-1000 hours of operation (which in turn depends on your operating conditions: if making more particles, clean more often). If higher settings are needed to maintain a spark flow, it is possible that the reactor needs to be cleaned to prevent spark obstruction which can possibly lead to spark misfiring within the reactor system. If the pressure within the system increases even with the valves open, it is possible that the filter is full and needs to be replaced. Refer to the [[VSP-G1_Operating_and_maintenance_guide|operating and maintenance guide]] for more information.
 
=== Maintenance Guide ===
The system requires no regular maintenance by the end user, but cleaning the reactor periodically is recommended, especially after a change of electrode material. To prevent exposure to nanoparticles and to ensure safe and efficient cleaning, guidelines are set out in the [[VSP-G1 Maintenance protocols|Maintenance guide]]. The procedure to replace electrodes is also discussed there.


=== Safe handling and storage ===
=== Safe handling and storage ===
<mark>content needed</mark>
The VSP-G1 is a scientific instrument, and must be handled with care. This section provides general guidelines on how to safely handle and store the system. Detailed methods are described in the [[VSP-G1_Operating_and_maintenance_guide | operating and maintenance guide]].
The VSP-G1 is a scientific instrument, and must be handled with care. This section provides general guidelines on how to safely handle and store the system. Detailed methods are described in section [[VSP-G1_User_Manual#Maintenance protocols|Maintenance protocols]].
{{Warning| [[VSP-G1_Operating_and_maintenance_guide#Flushing_the_system|Flush the system]] | before opening the system.}}


==== Disassembly ====
==== Disassembly ====
The base unit with reactor installed has a combined weight of 22kg. For general handling, e.g. when assembling a setup, the reactor is conveniently be transported separately from the base unit. A mount plate / transport tool is provided for this purpose. See [[VSP-G1_Maintenance_protocolsl#Unmounting the reactor assembly|Unmounting the reactor assembly]].
The base unit with reactor installed has a combined weight of 22kg. For general handling, e.g. when assembling a setup, the reactor is conveniently be transported separately from the base unit. A mount plate / transport tool is provided for this purpose. See [[VSP-G1_Operating_and_maintenance_guide#Unmounting_the_reactor_assembly|unmounting the reactor assembly of the operating and maintenance guide]].


==== General disposal guidelines ====
==== General disposal guidelines ====
===== Disposal of nanomaterials =====
===== Disposal of nanomaterials =====
Label and dispose cleaning materials contaminated with nanomaterials according to local rules and regulations. See also [[VSP-G1_Maintenance_protocolsl#Filter replacement|Filter replacement]] and [[VSP-G1_Maintenance_protocolsl#Disassembling and cleaning the reactor|Disassembling and cleaning the reactor]].
Label and dispose cleaning materials contaminated with nanomaterials according to local rules and regulations. See also [[VSP-G1_Maintenance_protocolsl#Filter replacement|Filter replacement]] and [[VSP-G1_Operating_and_maintenance_guide#Disassembling_and_cleaning_the_reactor|disassembling and cleaning the reactor]].


=====Retiring the VSP-G1=====
=====Retiring the VSP-G1=====
Line 371: Line 376:


==== Transportation, packaging and storage ====
==== Transportation, packaging and storage ====
VSParticle recommends using the original packaging for transportation and shipping. Be sure to [[VSP-G1_Maintenance_protocols#Disassembling and cleaning the reactor|clean the reactor]] before shipping the reactor. The system should also be cleaned if it is not to be used for a longer period of time.
VSParticle recommends using the original packaging for transportation and shipping. Be sure to [[VSP-G1_Operating_and_maintenance_guide#Disassembling_and_cleaning_the_reactor|clean the reactor]] before shipping the reactor. The system should also be cleaned if it is not to be used for a longer period of time.


When shipping the system to VSParticle or one of it's representatives for any reason, return authorization (RA) is required. Contact your distributor. For systems supported by VSParticle, an RA form can be provided by [mailto:support@vsparticle.org?Subject=VSP-G1%20RA%20Request VSParticle]. <mark>Vincent, graag aanvulling</mark>
When shipping the system to VSParticle or one of it's representatives for any reason, return authorization (RA) is required. Contact your distributor. For systems supported by VSParticle, an RA form can be requested by [mailto:support@vsparticle.org?Subject=VSP-G1%20RA%20Request sending an e-mail to support@vsparticle.com].


== Troubleshooting ==
== Troubleshooting ==


There are several possible issues that may arise during normal operation of the VSP-G1. A more thorough description of all error codes and ways to resolve these are listed in the [[VSP-G1 Troubleshooting guide|Troubleshooting guide]].
There are several possible issues that may arise during normal operation of the VSP-G1. The most common scenarios and errors are listed below.  


=== Unstable operation ===
In some scenarios, unstable operation of the spark (wild fluctuations of the gap voltage, intermittent sparking and/or the system is slow to reach its set point) may be encountered. There are several causes for such instability to occur:
* Glow mode is active. In this case, a weak arc is observed between both electrodes and the gap voltage drops suddenly. Please refer to the [[VSP-G1_User_Manual#Glow_mode | glow mode section below]].
* The carrier gas selected on the system differs from the carrier gas being used. Symptoms include wild fluctuations around the intended gap voltage set point and the motor in the system can be heard adjusting the gap all the time. Alternatively, the system may be very slow to reach the intended set point. To correct the issue, ensure the correct carrier gas is selected by following the instructions in [[VSP-G1_Operating_and_maintenance_guide#Changing_carrier_gas|the operating and maintenance guide]].
* Oxidation of the electrode material has occurred. This can result in fluctuations and/or intermittent sparking until the oxidised material has been ablated. To resolve this issue more quickly, increase the spark frequency by increasing the charging current set point and optionally the gap voltage (to avoid glow mode).


=== Glow mode ===
=== Glow mode ===
Line 388: Line 398:
|-
|-
| Interlock err. 7 || Motor controller failure || The motor controller reports a failure. This can be due to:
| Interlock err. 7 || Motor controller failure || The motor controller reports a failure. This can be due to:
* A blocked motor. Follow the instructions described on the [[VSP-G1 Maintenance protocols|Maintenance protocols page]] to dismount and disassemble the reactor. Then, verify that the electrode is free to move by manually rotating the gear located at the underside of the left end cap. Assemble and remount the reactor ensuring some space exists between both electrodes and power on the VSP-G1 to try again.
* A blocked motor. Follow the instructions described in the [[VSP-G1 Operating and maintenance guide|VSP-G1 Operating and maintenance guide]] to dismount and disassemble the reactor. Then, verify that the electrode is free to move by manually rotating the gear located at the underside of the left end cap. Assemble and remount the reactor ensuring some space exists between both electrodes and power on the VSP-G1 to try again.
* Motor controller communication interference or failure. Press both dials to clear the error and try again. Optionally, switch the machine off and on again. If the error persists, contact VSParticle.
* Motor controller communication interference or failure. Press both dials to clear the error and try again. Optionally, switch the machine off and on again. If the error persists, contact VSParticle.
* Motor controller failure. Press both dials to clear the error and/or switch the device off and on again. If the error persists, contact VSParticle.
* Motor controller failure. Press both dials to clear the error and/or switch the device off and on again. If the error persists, contact VSParticle.
Line 396: Line 406:
| Interlock Error 9 || Open interlock switch || One or more interlock switches are in the open state. Verify both reactor mounting pins are properly secured and that the reactor is properly mounted on the base unit. Press both dials to clear the error and try again.  
| Interlock Error 9 || Open interlock switch || One or more interlock switches are in the open state. Verify both reactor mounting pins are properly secured and that the reactor is properly mounted on the base unit. Press both dials to clear the error and try again.  
|}
|}
If you encounter any interlock error other than the ones listed above, please try to clear the interlock by pressing and releasing both dials and/or switch the VSP-G1 off and on again. If the error persists, please contact VSParticle via [mailto:support@vsparticle.com?Subject=VSP-G1%20Interlock%20Error support@vsparticle.com].


=== Electrode alignment ===
=== Electrode alignment ===
Line 405: Line 417:


When the electrode alignment sequence fails, this is displayed as ''Interlock err. 7''. Refer to the interlock error section above for possible resolutions.
When the electrode alignment sequence fails, this is displayed as ''Interlock err. 7''. Refer to the interlock error section above for possible resolutions.
=== Further troubleshooting ===
Refer to the [[VSP-G1 Troubleshooting guide|Troubleshooting guide]] for further troubleshooting your VSP-G1.


=== When to contact VSParticle ===
=== When to contact VSParticle ===
Line 420: Line 429:
'''Specifications of replaceable parts that can be obtained by other suppliers:'''<br>
'''Specifications of replaceable parts that can be obtained by other suppliers:'''<br>
''Fuses'' - The fuse holder accepts 5x20mm T2.5A/250V fuses.<br>
''Fuses'' - The fuse holder accepts 5x20mm T2.5A/250V fuses.<br>
''Power cable'' - IEC-<span style="background:#FF00FF">content needed</span>.<br>
''Power cable'' - IEC-60320 compliant with C13 connector,<br>
''RS-232 cable'' - any standard RS-232 cable is suitable.
''RS-232 cable'' - any standard (straight) RS-232 cable is suitable.


== Appendices ==
== Appendices ==
Line 430: Line 439:
Additional information about the VSP-G1 operating window, electrical specifications and output can be found on the [[VSP-G1 Technical Specifications|Technical Specifications]] page.
Additional information about the VSP-G1 operating window, electrical specifications and output can be found on the [[VSP-G1 Technical Specifications|Technical Specifications]] page.


=== Protocol index ===  
<!-- === Protocol index ===  
This section has been commented out. Use edit to view old content.
This section has been commented out. Use edit to view old content.
<!-- <mark>set this up as a table with the relevant location within the user manual (in case people decide to print out the manual... the location should be hyperlinked)</mark><br>
<mark>set this up as a table with the relevant location within the user manual (in case people decide to print out the manual... the location should be hyperlinked)</mark><br>
<mark>edit content</mark>
<mark>edit content</mark>


Line 473: Line 482:
-->
-->


=== Residual risks ===
<!--- === Residual risks ===
This section has been commented out. Use edit to view the original table.
This section has been commented out. Use edit to view the original table.
<!--
 
{| class="table table-bordered table-striped"
{| class="table table-bordered table-striped"
|-
|-
Line 483: Line 492:
!Mitigation of Hazard
!Mitigation of Hazard
|-
|-
|Setup ||Nanoparticle exposure ||Blind caps are not delivered already installed. If they are installed improperly, they might result in nanoparticle exposure. ||User action: installing blind caps to the appropriate places. User manual: warning about improper installation of blind caps <mark>(5.1)</mark>; leaktightness protocol <mark>(5.2)</mark>.
|Setup ||Nanoparticle exposure ||Blind caps are not delivered already installed. If they are installed improperly, they might result in nanoparticle exposure. ||User action: installing blind caps to the appropriate places. User manual: warning about improper installation of blind caps [[VSP-G1_User_Manual#Operator_exposure|Operator exposure]]; [[VSP-G1_User_Manual#Checking_for_leaks|leaktightness protocol]] .
|-
|-
|Setup ||High voltage exposure ||Failure to explicitly ground the unit through the protective earth terminal can result in exposure to lethal voltages if the main connection provides no or bad grounding protection. ||User action: grounding the unit through the protective earth terminal. User manual: information about grounding the unit <mark>(4.4)</mark>; warning about not grounding the unit <mark>(4.4)</mark>.
|Setup ||High voltage exposure ||Failure to explicitly ground the unit through the protective earth terminal can result in exposure to lethal voltages if the main connection provides no or bad grounding protection. ||User action: grounding the unit through the protective earth terminal. User manual: information and warning about grounding the unit about not grounding the unit [[VSP-G1_User_Manual#Grounding_the_system|Grounding the system]].
|-
|-
|Setup||Nanoparticle exposure||Not carrying out the leaktightness protocol can lead to possible nanoparticle exposure if there is a leak.||User action: leaktightness protocol. User manual: leaktightness protocol <mark>(5.2)</mark>; warning about nanoparticle exposure <mark>(5.1)</mark>.
|Setup||Nanoparticle exposure||Not carrying out the leaktightness protocol can lead to possible nanoparticle exposure if there is a leak.||User action: leaktightness protocol. User manual: leaktightness protocol [[VSP-G1_User_Manual#Checking_for_leaks|leaktightness protocol]]; warning about nanoparticle exposure [[VSP-G1_User_Manual#Operator_exposure|Operator exposure]].
|-
|-
| Setup; Operation||Nanoparticle exposure, explosion risk||Overpressurization during leaktightness protocol can lead to explosion of the viewport and compromising the gaskets, leading to possible nanoparticle exposure.||User action: ensuring that there is a pressure sensor connected to the setup and it is functioning. User manual: warning against overpressure <mark>(5.1)</mark>. Marking: pressure limit information label on base unit.
| Setup; Operation||Nanoparticle exposure, explosion risk||Overpressurization during leaktightness protocol can lead to explosion of the viewport and compromising the gaskets, leading to possible nanoparticle exposure.||User action: ensuring that there is a pressure sensor connected to the setup and it is functioning. User manual: warning against overpressure [[VSP-G1_User_Manual#Other residual risks|Other residual risks]], information on pressure monitoring [[VSP-G1_User_Manual#Pressure_monitoring|Pressure monitoring]]. Marking: pressure limit information label on base unit.
|-
|-
| Operation||Reactor damage||Using the wrong gas settings or reverse flow of gas can reduce the lifetime of the reactor.||User action: ensuring that correct gas settings are used and prevention of reverse gas flow. User manual: warning against use of wrong settings or reverse flow <mark>(WHERE?)</mark>. <mark>[content needed]</mark>
| Operation||Reactor damage||Using the wrong gas settings or reverse flow of gas can reduce the lifetime of the reactor.||User action: ensuring that correct gas settings are used and prevention of reverse gas flow. User manual: warning against use of wrong settings or reverse flow [[VSP-G1_User_Manual#Flow_configurations|Flow configurations]].
|-
|-
|Operation||Reactor damage||There is a risk of damage to the setup if a rigid gas connection is used on the inlet only valve, due to the movement of the valve during the homing sequence.||Design feature: Display asks for confirmation before initiating the homing sequence. User manual: warning against the use of a rigid gas connection on the inlet only valve <mark>(4.3)</mark>.
|Operation||Reactor damage||There is a risk of damage to the setup if a rigid gas connection is used on the inlet only valve, due to the movement of the valve during the homing sequence.||Design feature: Display asks for confirmation before initiating the homing sequence. User manual: warning against the use of a rigid gas connection on the inlet only valve [[VSP-G1_User_Manual#PRocess_connections|Process connections]].
|-
|-
|Operation||Fire risk||There is a fire risk when working with very pure carrier gas (oxygen free) and metal nanoparticles.||User action: ensure (introduction of) small amount of oxygen in the system before opening. User manual: warning of fire risk without flushing the system; list of ways to safely open the system <mark>(5.1)</mark>.
|Operation||Fire risk||There is a fire risk when working with very pure carrier gas (oxygen free) and metal nanoparticles (pyrophoricity).||User action: ensure (introduction of) small amount of oxygen in the system before opening. User manual: warning of fire risk without flushing the system; list of ways to safely open the system [[VSP-G1_User_Manual#Operator_exposure|Operator exposure]].
|-
|-
|Operation||Nanoparticle exposure, gas exposure||Opening the viewport or blind caps during operation can expose the user to nanoparticles and carrier gas.||Design feature: an additional screw on the viewport clamp makes it more difficult for the operator to deliberately remove the viewport. User manual: warning against removal of viewport and blind caps during operation <mark>(5.1)</mark>. Marking: warning label on the viewport.  
|Operation||Nanoparticle exposure, gas exposure||Opening the viewport or any process connections during/after operation can expose the user to nanoparticles and carrier gas.||Design feature: an additional screw on the viewport clamp makes it more difficult for the operator to deliberately remove the viewport. User manual: warning against removal of viewport and blind caps during operation [[VSP-G1_Maintenance_protocols#Cleaning_the_viewport|Cleaning the viewport]]. Marking: warning label on the viewport.  
|-
|-
|Operation||High voltage exposure||There is a risk of exposure to lethal voltages if the topcap of both endcaps are opened.||Design feature: special tool is needed to unscrew the top plate of the endcaps. User manual: warning against opening the top of the endcaps <mark>(3.1.1)</mark>. Marking: high voltage warning labels on both topcaps on the reactor endcaps.
|Operation||High voltage exposure||There is a risk of exposure to lethal voltages if the topcap of both endcaps are opened.||Design feature: tool is needed to unscrew the top plate of the endcaps, and a tamper-proof sticker must be removed. User manual: warning against opening the top of the endcaps [[VSP-G1_User_Manual#Setting_up_the_base_unit|Setting up the base uit]]. Marking: high voltage warning labels on both topcaps on the reactor endcaps.
|-
|-
|Operation||High voltage exposure, gas exposure, nanoparticle exposure||There is a risk of exposure to lethal voltages, carrier gas and nanoparticles if the spark is enabled (remote/local) when the setup is opened without proper shutdown.||Design feature: unit display will indicate that the unit is being operated locally. User can push a button on the unit to cancel remote operation. User manual: warning against working with the unit while it is being operated remotely <mark>(4.5)</mark>.
|Operation||High voltage exposure, gas exposure, nanoparticle exposure||There is a risk of exposure to lethal voltages, carrier gas and nanoparticles if the spark is enabled (remote/local) when the setup is opened without proper shutdown.||Design feature: unit display will indicate that the unit is being operated locally. User can push a button on the unit to cancel remote operation. User manual: warning against working with the unit while it is being operated remotely <mark>(4.5)</mark> Update: Remote and local control work simultaneously. Interlocks are used to guard against exposure. After power down system needs manual start before remote can work.
|-
|-
|Operation||Reactor damage||There is a risk of damage and/or injury when using heated carrier gas. (Environmental temperatures are specified, but not gas specs).||User manual: warning against use of heated gas <mark>(4.3)</mark>.
|Operation||Reactor damage||There is a risk of damage and/or injury when using heated carrier gas. (Environmental temperatures are specified, but not gas specs).||User manual: warning against use of heated gas [[VSP-G1_User_Manual#Process_connections|Process connections]].
|-
|-
|Cleaning and Maintenance; Storage and Transport||Nanoparticle exposure||Risk of nanoparticle exposure when: reactor is opened outside safe environment; the system is not properly vented when the gas is disconnected; reactor is dropped.||Design feature: use of the transport plate provides stability to the reactor and its clamps, helping reduce risk of opening the reactor by accident while transporting it to a safe environment. User manual: proper procedures for shutting down the reactor and preparing it for disassembly and transport to a safe environment <mark>(4.5, 5.1, 6.2)</mark>. Marking: Warning to read and understand the User Manual before use.
|Cleaning and Maintenance; Storage and Transport||Nanoparticle exposure||Risk of nanoparticle exposure when: reactor is opened outside safe environment; the system is not properly vented when the gas is disconnected; reactor is dropped.||Design feature: use of the transport plate provides stability to the reactor and its clamps, helping reduce risk of opening the reactor by accident while transporting it to a safe environment. User manual: proper procedures for shutting down the reactor and preparing it for disassembly and transport to a safe environment [[VSP-G1_User_Manual#System_care|System care]], operating protocols. Marking: Warning to read and understand the User Manual before use.
|-
|-
|Cleaning and Maintenance; Storage and Transport||Reactor damage||Risk of reactor damage if it is accidentally dropped.||Design feature: use of the transport plate helps reduce risk of damage to the reactor in the case that it is dropped because it keeps all the potentially detachable components together. User manual: warning to use the transport plate whenever handling the detached reactor <mark>(3.1.2)</mark>.
|Cleaning and Maintenance; Storage and Transport||Reactor damage||Risk of reactor damage if it is accidentally dropped.||Design feature: use of the transport plate helps reduce risk of damage to the reactor in the case that it is dropped because it keeps all the potentially detachable components together. User manual: warning to use the transport plate whenever handling the detached reactor [[VSP-G1_User_Manual#System_care|System care]].
|-
|-
|Cleaning and Maintenance||Reactor damage||Using the wrong solvent and/or solvent film on the insulator can result in a short circuit and a broken reactor.||User manual: proper procedure for cleaning the reactor, including appropriate cleaning supplies <mark>(6.2)</mark>. Marking: Warning to read and understand the User Manual before use.
|Cleaning and Maintenance||Reactor damage||Using the wrong solvent and/or solvent film on the insulator can result in a short circuit and a broken reactor.||User manual: proper procedure for cleaning the reactor, including appropriate cleaning supplies [[VSP-G1_Maintenance_protocols#Disassembling_and_cleaning_the_reactor|Disassembling and cleaning the reactor]]. Marking: Warning to read and understand the User Manual before use.
|-
|-
|Cleaning and Maintenance||High voltage exposure, nanoparticle exposure||Exchanging the easy-open clamp with the viewport clamp (which requires a tool) poses a risk because the viewport can then be opened easily during operating resulting in high voltage and nanoparticle exposure.||User manual: describes the risk of not having a viewport in place <mark>(3.1.1)</mark>; describes the proper clamps to use for the viewport <mark>(3.1.1)</mark>. Marking: warning labels on the viewport warn against removal of the viewport while the spark generator is running; label on clamp labeling it as the viewport clamp.
|Cleaning and Maintenance||High voltage exposure, nanoparticle exposure||Exchanging the easy-open clamp with the viewport clamp (which requires a tool) poses a risk because the viewport can then be opened easily during operating resulting in high voltage and nanoparticle exposure.||User manual: describes the risk of not having a viewport in place ; describes the proper clamps to use for the viewport [[VSP-G1_Maintenance_protocols#Cleaning_the_viewport|Cleaning the viewport]]. Marking: warning labels on the viewport warn against removal of the viewport while the spark generator is running; label on clamp labeling it as the viewport clamp.
|-
|-
|Storage and Transport||Nanoparticle exposure||There is a risk of nanoparticle exposure when the unit is shipped or stored without cleaning it.||User manual: protocol for preparing the unit for storage and transport <mark>(6.3 and 6.4)</mark>. <mark>[content for 6.3 and 6.4 needed]</mark>
|Storage and Transport||Nanoparticle exposure||There is a risk of nanoparticle exposure when the unit is shipped or stored without cleaning it.||User manual: protocol for preparing the unit for storage and transport [[VSP-G1_User_Manual#System_care|System care]]
|}-->
|}
-->


== EC Declaration of Conformity ==
<!--== EC Declaration of Conformity ==
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']<br>
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']<br>
<span style="background:#FFFF00">''version for user documentation does not need to show signature nor serial number''</span>
<span style="background:#FFFF00">''version for user documentation does not need to show signature nor serial number''</span>


== Warranty ==
== Warranty ==
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']
[https://docs.vsparticle.com/w/index.php/G1_User_Manual ''top'']-->
 
== Warnings ==
<span style="background:#FFFF00"> Note: I left all of the warning labels from the original draft of the user manual. Not all have been inserted into the document itself. These can be edited for an "Important Safety Guidelines" document to be shipped with the new VSP-G1 unit. - Kate</span>
 
All warnings hidden for now, please use edit to view the commented section.
<!--
{{Warning|| Explosion risk: Do not exceed +0.4 bar. Do not turn on the spark below -0.4 bar.
}}
 
{{Warning|| Fire risk: The VSP-G1 is not intended for use with reactive gases, in particular gases containing more than a few percent oxygen, such as air.
}}
 
{{Warning|| Reactor damage risk: Do not use heated gases as carrier gases.
}}
 
{{Warning||Intentional misuse of the VSP-G1 in a manner not specified by VSParticle may impair the existing protective features.}}
 
{{Warning||Gas connections and other user-provided functions are not controlled by the interlock.}}
 
{{Warning||Do not use a rigid gas connection on the gas inlet valve. You can damage the reactor if the valve is not free to move during the homing sequence.}}
 
 
{{Warning||Ensure that your pressure gauge is working properly before operating the system.}}
 
{{Warning||The optical filter on the viewport is designed to protect your eyes against the intense UV light of the spark. Do not remove the filter when running the spark generator.}}
 
{{Warning||Failure to follow protocol or lack of external monitoring tools could mean the total output of the system&mdash;that is, up to 100 mg/h of nanoparticles&mdash;is released continuously to the operator.}}
 
{{Warning||Do not handle or open the system while particles are being generated and while gas is flowing.}}
 
{{Warning||VSParticle does not recommend the use of air or other oxygen-rich gases for nanoparticle production in the VSP-G1 unit.}}
 
{{Warning||Do not use the power supply in the topcaps as an outlet.}}
 
 
{{Warning||When conducting activities requiring an open reactor, ensure that the power is off, and that the reactor assembly has been moved to a safe environment before it is opened, to avoid possible nanoparticle contamination. If pressure has not been equalised, there's also a chance of explosion.}}
 
{{Warning||The viewport is specially coated to screen electromagnetic radiation. Do not scratch the coating (see <span style="background:#00FF00">maintenance</span>). Do not replace the viewport with a different type.}}
{{Warning||Do not remove the filter on the viewport, which provides eye protection. If the viewport needs to be cleaned, make sure the spark generator is turned off and the reactor has been relocated to a safe environment.}}
 
{{Warning||Improper installation of blind caps on the unused valves can result in possible user exposure to nanoparticles and carrier gas.}}
 
{{Warning|High voltage risk:| do not open the base unit. There are no user serviceable parts inside.}}
 
{{Warning||Do not replace the detachable mains supply cords with inadequately rated cords.}}
 
{{Warning||If the power cable provided with the VSP-G1 unit is not functional, please use a power cable with an integrated protective earthing terminal. Otherwise, take measures to ensure that the system is earthed. (See <mark>"Grounding the system"</mark> below.)}}
 
{{Warning|| Foreseeable misuse: xx}}
 
{{Warning||Ensure that all surfaces are dry before beginning assembly of the G1 unit.}}
 
{{Warning||Under no cirumstances mount the endcaps without a sealed reactor, such as the one provided. The reactor should be conductive, preferably metal.}}
 
{{Warning||Always screw down '''both''' alignment pins.}}
 
{{Warning||The mains supply has protective earthing, but always ground the system using the external functional earth terminal, which provides additional protection in case the first set of protections fail to work.}}
 
{{Warning||The <mark>leak tightness protocol</mark> is an essential part of the operating procedure before initiating particle production.}}
 
{{Warning||Please read the entire procedure before using the VSP-G1!}}
 
 
{{Warning||The spark generator should not be running during this procedure or you risk unnecessary exposure to nanoparticles!}}
 
{{Warning||Ensure that you are wearing safety equipment including gloves when working with an open reactor.}}
 
{{Warning||Do not apply pressure when wiping the surface.}}
 
{{Warning||Ensure that all surfaces are dry before operating the spark generator!}}
 
{{Warning||When reassembling, make sure to retract the electrode sufficiently to avoid collision with the right electrode.}}-->

Latest revision as of 14:30, 30 March 2023

This is the comprehensive User Manual for the VSParticle Generator One Spark Generator (VSP-G1). The Quick Start Guide is available as a quick reference for standard operation of the VSP-G1 unit.

Company VSParticle B.V.
VSParticle Generator One
Address Oostsingel 209
2612 HL
Delft
The Netherlands
Website vsparticle.com
Support e-mail support@vsparticle.com

Introduction

The User Manual provides instructions for the set-up, installation, operation and maintenance of a VSParticle Generator 1 ("VSP-G1") unit.

Alert Warning: Before operating the VSP-G1 unit, please read—and understand—this manual.

Wherever a caution CAUTION symbol is affixed to the machinery, there are explanatory texts located within the manual (see this section of the detailed system description for a guide to all symbols found on the VSP-G1. ). Other important safety information located within the manual are also labelled with the caution CAUTION symbol.

You will also find conformity and warranty information at the very end of this user manual.

A Quick Start Guide is also included for faster set-up as part of the standard operating procedure. For more information regarding safety, more complex operations and troubleshooting and maintenance, refer to this manual first!

If you have any questions or concerns regarding the VSP-G1 unit you have purchased or any information within the provided user documentation, please contact us via the contact details at the top of this page.

VSParticle Generator One

The VSP-G1 is designed for use by researchers who want to produce inorganic nanoparticles from conductive materials for applications in fields ranging from cluster research to materials science. The VSP-G1 is a desktop spark generator that uses spark ablation technology to produce aerosol nanoparticles after a high voltage spark evaporates part of two metallic electrodes. Its design is aimed at making the production of nanoparticles as simple as possible, while giving you the flexibility to customise the system by having control over:

  • particle size (1 atom to 20 nm)
  • ablation rate (~0.01-100 mg/h)
  • carrier gas
  • electrode material
  • gas flow rate (1-30 l/min)


It is also possible to customise the reactor by changing configurations from through flow (the standard configuration offered with the basic VSP-G1 unit) to cross flow, co-flow and combined configurations. The portability of the system also allows it to be integrated into a larger system for more complex operating procedures.

The VSP-G1 is designed as a continuous source of aerosolized nanoparticles for use in a controlled process system. This controlled process system typically comprises a gas source, a gas-tight system of unit operations, valves and tubing, and a gas exhaust fitted with one or more particle filters. Typical conditions in the process system are room temperature and ambient pressure. The system can be run at slight over- or under-pressure.

The closed system provides a safer production of nanoparticles and can easily be dismantled for safe transport of produced nanoparticles and for system cleaning and maintenance, which can be performed following a simple and fast protocol. The closed system design also allows you to produce high purity nanoparticles without any polluting by-products. The spark ablation process eliminates the use of any chemicals, surfactants, precursors or polymers. Without the need to rely on external materials to stabilise the produced particles, you have more flexibility to choose a suitable deposition system for the aerosol nanoparticles. The three physical principles to inmobilize the nanoparticles on a support are diffusion, impaction and electrostatic precipitation. If you have any questions or need help with the selection and/or development of a system that best fits your needs, please contact us.

For more information about the spark ablation process and possible applications, refer to the Background information section.

System and operator safety

top

Notification: To view a list and description of all warning labels found on the VSP-G1 unit, go to the Guide to symbols found on the VSP-G1

The risks associated with a spark generator can be divided into two categories: those associated with the machine itself, and those associated with nanoparticles. Machine-related risks are more widely regulated by international norms, such as the European conformity system required for marking certain products with the CE symbol. However, synthesized nanomaterials are relatively new and their quantum effects are still being studied, thus there are less standardized measures in place. Until more is known about nanoparticles, we apply the Precautionary Principle, treating them as potential highly toxic materials. The safety measures we employ for working with, storing and transporting nanoparticles have been drawn primarily from those designed for fine particles.

Safety of machinery

The VSP-G1 spark generator comprises general electronics (<50 V), gas connections and high voltage (6 kV), as well as moving parts. Several safety mechanisms mitigate the risk for damages and incidents related to potential equipment failure, but primary responsibility for correct and safe operation lies with you as the operator of the system. The VSP-G1 is intended for use by qualified personnel only.

Operator safety actions

You are required to ensure that:

  • the reactor is correctly mounted (see setting up);
  • there are no objects near the electrodes;
  • process connections are correctly made (see process connections);
  • the pressure in the system is limited to 1±0.2 bar absolute (see gas pressure monitoring);
  • all, if any, external safety systems are functional (where applicable); and
  • personal protective equipment are being used correctly (gloves, glasses, and FFP3/P3 certified filter mask).

Existing protective features

Alert Warning: Intentional misuse of the VSP-G1 in a manner not specified by VSParticle may impair the existing protective features.

The VSP-G1 is equiped with an interlock system. When a failure is detected, the interlock system cuts power to the high voltage supply and the motor, shutting down particle production. There are three types of interlock.

1. Internal Interlock
During start-up and while running, the VSP-G1 the control system checks if internal systems are working properly. If any check fails an internal interlock is triggered. When the interlock cuts power, you must first fix the conditions that caused the interlock to activate before clearing the interlock state. Press and depress both dials simultaneously to clear the interlock.

2. Reactor interlock

The mounting pin is used both to secure the reactor assembly, and to close the interlock protection.

The VSP-G1 detects whether the reactor endcaps are properly mounted, using two mounting pins (see image). You must ensure that the reactor itself is correctly assembled. In short, this means that the reactor is leak tight (see the operating and maintenance guide) and that the gas inlet valve and electrode tips are free to move with a minimum clearance of 10 mm from any foreign objects (see the section on process connections and the section on flow configurations).

Alert Warning: Gas connections and other user-provided functions are not controlled by the interlock. An optional external interlock is available for the VSP-G1.

3. External interlock (option)
An external interlock switch can be provided for incorporation in an external safety system. To allow operation, two pins of the interlock connector must be connected (shorted). Breaking the connection switches the system into interlock mode.

Other residual risks

Pressure. Pressure within the closed system poses a serious risk when improperly handled. Increased pressure can be caused by unintentional release of gas at high pressure, the closing of output valve while the input valve is open, clogged filters, spark obstruction, or a full deposition system. Exceeding the maximum pressure can lead to failure of the gaskets and seals, and the viewport could explode.

Non-ionizing radiation. The spark plasma emits a bright light in the visible and UV range. A viewport with a removable optical filter is included to allow visual inspection of the electrodes. Additionally, electromagnetic radiation (radio frequency) may be emitted.


Safety of nanoparticles

Because our spark generators can in principle be used to produce nanoparticles from any metal, we start from the assumption that these nanoparticles pose potential risks to the health of operators, consumers and to the environment. This manual addresses operator exposure. Consumer exposure is determined by the application in which the nanomaterials are to be used and should be evaluated on a case-by-case basis by the user. Environmental considerations should also be addressed by the operators with a life cycle analysis.

The main channels of operator exposure to nanoparticles are through (skin) absorption, ingestion and inhalation. The aerosol nature of the produced nanoparticles by the VSP-G1 system means exposure through inhalation is the main risk. The respiratory system of the human body has natural defenses in place for airborne particles such as dust, viruses and bacteria. However, nanoparticles are able to bypass most of these defenses due to their small size. For this reason, nanoparticles are treated in the same way as fine particles. More information on health and safety of nanoparticles can be found through the links at the end of this section.

Operator exposure

While the VSP-G1 unit is designed to be a closed system to maximize safety when working with nanoparticles, there will be situations in which it will be necessary to open the system, including routine maintenance activities, replacement of electrodes and filters, and sample extraction. The list below also includes some possible situations in which the operator may be inadvertently exposed to nanoparticles and a brief summary of the necessary precautions to prevent, wherever possible, such situations.

Exposure risk situation Operator precautions
Gas phase exposure An aerosol system must be leak tight from source to deposition (see section 2.1 of the operating and maintenance guide).
Filtration The gas exhaust should be treated with a suitable HEPA or ULPA filter. VSParticle recommends using a large capacity HEPA pre-filter, followed by an ULPA end filter.
Dealing with pyrophoric materials Metal nanoparticles can be extremely reactive, and can be pyrophoric (spontaneous ignition upon exposure to air).Controlled air-exposure, however, is needed before opening the system in order to passivate any reactive materials (see controlled-air exposure protocol).
Product handling When the system must be opened to insert or remove a substrate, methods as described in the maintenance protocols section must be followed to reduce the risk of nanoparticle exposure.
Maintenance Nanoparticles will collect on the surfaces inside the reactor and downstream piping. Cleaning of these surfaces presents the largest exposure risk to operators, unless they can be performed in a safe environment (e.g. a glovebox) (see system care and the operating and maintenance guide for the relevant maintenance protocols).


There is also a separate protocol for dismantling and sealing used filters for disposal/recycling, which is largely dependent on your particular filtration system (see the section on filter replacement in the operating and maintenance guide).

Links to more information on nanoparticles and health

System description

top

What's in the box

Congratulations on your purchase of the VSP-G1!

In the box you will find the following:

You will also find a Quick Start Guide for daily operation and a reference card for working with nanoparticles in the lab. The reference card is also available online as part of the Quick Start Guide. If there's anything missing or looks damaged, contact us immediately at support@vsparticle.com or using the contact details at the top of this page! A damaged power supply cable may temporarily be replaced by a similar cable, but ensure it has a functional protective earthing terminal.

Detailed system description

A visual reference and detailed description of the components that are part of the VSP-G1 system is provided on the Detailed System Description page. This includes a full reference of the warning labels and symbols found on the machine.

Setting up

top

Setting up the base unit

Alert Warning: When conducting activities requiring an open reactor, ensure that:
  • The power is off and the power supply has been disconnected;
  • There is no gas flowing through the system;
  • You are wearing safety equipment, including gloves;
  • The reactor assembly has been moved to a safe environment before it is opened (unless handling for the first time); and that
  • The internal pressure has been equalised (unless handling for the first time).
Alert Warning: Do not open the base unit or the endcaps. There are no user serviceable parts inside.
Front view of the VSP-G1 base unit
Place the base unit on a flat surface, ensuring sufficient space around the sides and back of the unit for gas, power and communication connections.

Do not connect the system to a power supply until everything has been properly assembled.

Mounting the reactor assembly

Alert Warning: Under no circumstances mount the endcaps onto the base unit without a sealed reactor.

When shipped, the reactor assembly is attached to the transport plate in the packaging. The electrodes have already been installed. In the case you want to replace the electrodes, please see the electrode replacement protocol in the operating and maintenance guide.

Step Description Picture
1. Place the reactor assembly and transport plate close to the VSP-G1 base unit and unscrew and remove the alignment pins holding the reactor assembly to the transport plate.
2. Holding the endcaps in both hands, carefully transfer the reactor assembly to the mount plate located at the top of the base unit, aligning the markings on the bottom of the reactor assembly's endcaps and the mount plate.
3.

Lower the reactor assembly gently and allow it to slide into position.

4.

Use gentle pressure to ensure the reactor assembly is fully pressed down.

5.

Insert the alignment pins and tighten.

Checking the resistance between the reactor chamber and the ground point.

To confirm that the reactor assembly is correctly assembled, use a digital multimeter (DMM) to measure the resistance between the reactor chamber and the ground point on the back of the system. The resistance should be <0.1 Ohm (keep in mind that the leads in a typical handheld DMM can have a resistance as high as 0.5 Ohm).

Building a basic setup

Your particular deposition requirements and research needs will result in a system setup that is different from others. In all cases, however, the following requirements help ensure that the system works efficiently and safely:

  • Use correctly rated tools and components
  • Apply compartmentalisation of the reactor and the deposition system;
  • Use a gas pressure monitoring tool;
  • Use a safe exhaust (fume hood with HEPA+ filters, for example) for the outflowing gas; and
  • Perform regular leak checks.


A simple setup connecting the input gas source and flow system to the particle source, deposition system, filtration system and the ventilation system.

The following basic operating procedure is for the simple setup pictured below, which comprises several operating units: a gas source and flow system; a particle source (the VSP-G1 unit); a pressure gauge; a deposition system; and exhaust filters. To complete the setup, the units are connected with tubes, piping, fittings, valves and so on. The ball valves () in this setup help compartmentalize the overall system for leak testing and for safe disconnection. Each setup will be different, depending on your individual needs, but the concepts described here should generally be valid.

To reduce particle losses, please take the following guidelines into account:

  • Keep the tubing length between particle source and deposition system as short as possible to reduce diffusive losses.
  • Use conductive tubing (metal or conductive polymer (electrostatic dissipative tubing) in particle carrying lines to reduce electrostatic losses.
  • Do not use flow controllers (e.g. needle valves, mass flow controller) in particle lines, as particles will clog up the controller.
    • Usually the controller can be placed in another (particle free) line, or downstream of the deposition system, behind a filter.
    • If flow control is absolutely needed in a particle line, use an orifice.
  • Avoid constrictions that can function as a particle trap. I particular:
    • Use ball valves with a wide bore.
    • Avoid fibers and meshes.

Building a setup takes three steps:

  1. Design/draw/sketch your flow diagram, including sensors, valves and filters.
  2. Assemble the VSP-G1 (see Assembling the system
  3. Assemble the rest of the setup according to your flow diagram.
  4. Check the system for leak tightness and make adjustments if necessary. (See Checking for leaks.)

Process connections

Use only tubing and components rated for the operating conditions of the system. When using quick-connect fittings and other systems that do not require tools, ensure that the system is protected against accidental opening (e.g. clamps, latches and/or signs). Different technologies can be used to connect tubing and the other components. The process connections must be made according to the fitting vendor's specifications. The VSP-G1 comes equipped with Swagelok® fittings.

Alert Warning: Do not use a rigid gas connection on the gas inlet. Allow for 25mm movement of the reactor inlet tube. Damage may occur if the inlet tube is not free to move during the electrode alignment procedure.
Alert Warning: Important notes about carrier gases:
  • Do not use reactive gases, in particular gases containing more than a few percent oxygen, such as air.
  • The VSP-G1 is designed for use at room temperature. Do not use heated gases.


Compartmentalisation

Compartmentalisation of the setup.

For safety reasons, we recommend compartmentalisation of every unit within the setup that regularly opens up to the atmosphere to allow for easy dismantling in a safe manner. In the setup described here, we use two valves on both sides of the deposition chamber which allows us to easily seal off both the reactor and the rest of the setup. When the deposition chamber is removed, it can be safely transported to a safe working location, such as a fumehood or glovebox, in order to take out the sample. Meanwhile, the rest of the setup, which often contains a significant amount of nanomaterial, remains safely sealed.

Gas pressure monitoring

Alert Warning: Do not exceed 1.2 bar absolute. Do not turn on the spark below 0.8 bar.

The internal pressure should not exceed +1.2 bar or there is an increased risk of failure of the gaskets and/or seals in the reactor, causing leaks and risk of nanoparticle exposure. A build-up of pressure within the system might be an indicator of obstruction, either within the reactor, the deposition system or in the filtration system. Ideally, a pressure sensor at the reactor inlet is used to automatically switch off the gas supply. Stop spark production and gas flow immediately when pressure reaches +1.2 bar to determine the source of obstruction and resolve the issue. See System care for cleaning procedures.

Flow configurations

The VSP-G1 has multiple inlets and outlets, that allow the user to modify the flow profile around the spark gap. Configurations usually have a single outlet for the particles, but can have a single or multiple inlets. Possible reasons to modify the flow configuration are to change the dilution rate in the gap, or to apply a sheath or dilution gas. Particular attention must be taken if the hollow electrode is used as the only outlet: During the startup sequence, the VSP-G1 lets the electrodes touch, obstructing the outlet flow.

Alert Warning: Pressure may build up in the reactor if the hollow electrode is the only outlet, and one of the side inlets is used as an inlet. In such configurations, the user must ensure pressure cannot rise above 1.2 bar absolute.
Alert Warning: When using the hollow electrode as the only outlet, be sure not to mount a solid electrode tip.

Exhaust

The gas exhaust should be treated with a suitable HEPA or ULPA filter. VSParticle recommends using a large capacity HEPA pre-filter, followed by an ULPA end filter. Replace the pre-filter before the end-filter is filled. The filtered exhaust should not be vented directly into a workspace; vent the outflowing gas through a fume hood or other suitable (non-recirculating) ventilation exhaust.

Checking for leaks

Alert Warning: The leak tightness protocol is an essential part of the operating procedure before initiating particle production. Failure to test for leak tightness, improper installation/removal of blindcaps or lack of external monitoring tools could mean the total output of the system—that is, up to 100mg/h of nanoparticles—is released continuously to the operator.

Refer to the operating and maintenance guide for information on how to perform a leak test on the system.

Electronic connections

Grounding the system
The spark generator uses the ground from the mains plug for safety. Use only certified cables and properly wired wall outlets. The separate functional earth terminal on the back of the base unit (#21 in the base unit back diagram above) is internally connected to the mains connector ground, and should be independently connected to the protective ground in the case a recommended mains power supply plug is unavailable.

Alert Warning: Failure to sufficiently ground the VSP-G1 unit can result in exposure to lethal voltages.

Power supply connection
After assembling the base unit and reactor, plug in the provided mains cable into the mains inlet and connect the VSP-G1 to a mains outlet (110-230VAC).

Communication The VSP-G1 comprises a female 9-pin sub-D connector.

Pin Purpose
2 Tx (output)
3 Rx (input)
5 Common
Unassigned pins are not connected.

The VSP-G1 can be remotely controlled using RS232 over the sub-D connector. Please refer to the section 5.1.2 on remote controlled operation.


Operating the system

top

Alert Warning: Please read the entire Basic operating procedure before operating the VSP-G1!
Alert Warning: Do not handle or open the system while particles are being generated and while gas is flowing.

Controls

Manual control

The VSP-G1 can be controlled directly from the front panel using the two dials and the button on the left hand side. The two dials on the front panel can be pressed to select a function, such as changing a set point. In general, the dials can be used to change the value shown directly above it on the display by pressing the dial and then rotating it to change the associated value. Another press of the dial then locks or confirms the value. The button on the left hand side of the front panel is used solely to toggle sparking. Refer to the controls and indicators section of the detailed system description for additional information. This section further describes the key functions of the dials and button in various states of operation of the VSP-G1.

Starting the alignment sequence

When powering up the device, the message "Begin electrode alignment" appears on the display. Confirm no carrier gas is flowing and that the gas inlet is free to move, then confirm by pressing the right hand dial.

Changing the charging current and/or gap voltage set point

This can be done during sparking and when the "Ready to spark" message appears on the display. Press the dial underneath the value you wish to change and then rotate the dial to change the set point. Press the dial again or leave it untouched for 2 seconds to confirm the value.

Start/stop sparking

Sparking can only be initiated when the "Ready to spark" message is shown on the display. Press the button to start sparking. When sparking, another press of the button will return the system to the idle "Ready to spark" mode.

Accessing and navigating the menu

When the "Ready to spark" message is displayed, press and release both dials simultaneously to access the menu. By turning the right hand dial, you can browse the menu options. Press the right hand dial to confirm or alter a menu setting. When altering a setting, press the right hand dial again to confirm the value. Pressing the left hand dial exits the menu and returns the system to the "Ready to spark" mode.

Remote controlled operation

Alert Warning: Do not open the system while it is being operated remotely.

The VSP-G1 can be controlled remotely via the 9 pin D-SUB connector at the back of the system. The remote control structure allows for changing set points and reading the status of the device. It does not allow access to the settings menu and cannot initiate the electrode alignment procedure during start up. A full discussion of the command structure, functions and limitations can be found in the Remote control guide.

Basic operating procedure

An abridged form of this section is also available as a Quick Start Guide, including a themed version for printing.

Preparation

  1. Ensure that the system is leak tight.
  2. Power up all systems, includeing the VSP-G1 using the mains switch on the back panel near the power connector. Ensure the power cable is properly fitted and powered.
  3. When the "Start electrode alignment?" message appears on the display, press and release the right hand dial on the front panel to start the alignment sequence.
  4. Wait for the alignment procedure to complete. Several messages appear on the display during the sequence to indicate activity. The integrated LED on the button will blink during the entire sequence and stays on once completed.
  5. Ensure that the path to the exhaust is clear and all required valves are open.
  6. Start the carrier gas flow.
  7. Monitor the pressure to ensure it is within the required range (0.8-1.2 bar(a)).

Notification: Carrier gas selection: The VSP-G1 supports Ar and N2 as carrier gases. The system needs to be configured for the carrier gas that you use. Follow the instructions in the operating and maintenance guide to verify the right carrier gas setting is selected. If set incorrectly, the spark will be unstable or very slow in reaching the desired setpoints.

Start and stop nanoparticle production

The set points for charging current (frequency control) and gap voltage may be altered when the system displays the "Ready to spark" message and when the system is sparking. The set points are controlled by the two dials on the front panel. The left hand dial is associated with the gap voltage setpoint while the right hand dial controls the charging current. Higher voltage and current generally lead to larger particles.To change a setpoint:

  1. Press and release the corresponding dial. The unit of the associated value (e.g. kV or mA) will start blinking indicating the value is unlocked.
  2. Rotate the dial to set the desired value.
  3. Press and release the dial once more or wait 2 seconds to lock the value.

To start sparking, press and release the button on the left hand side of the front panel. The indicator LED around the button will change from solid to blinking red, indicating the machine is in operation. To stop sparking, press and release the button once more. This will return the system to the "Ready to spark" mode. It's possible to adjust the particle output by changing the voltage and current settings.

We recommend leaving the system to equilibrate/stabilise for 15-30 minutes, depending on your specific needs. This time allows the system to warm up, and the surface of the electrode tips to be refreshed and removing, if any, surface oxides.

Shutting down

  1. Make sure the system is in "Ready to spark" mode.
  2. Press and release both dials simultaneously to enter the menu. The option to shut down the system is selected upon reaching the menu.
  3. Press and release the right hand dial to confirm shutdown.
  4. Wait for the message "Please power off" to appear
  5. Set the mains switch near the power cable at the back of the system to the 0 position.

Opening the system

Four valves, two on each side of the deposition system have been closed to contain it for removal from the setup. When removing the deposition system, the contaminated surface that the operator can be exposed to is reduced to the volume between the valves.

Common reasons to open the system are to place or remove a sample/substrate, to clean the reactor, or to change electrodes. To reduce the risk of exposure when handling non-immobilized nanomaterials, it's recommended to open the system in a controlled working environment. If the deposition chamber is not equipped with sufficient protective measures, it can be convenient to move the deposition chamber itself to a fume-hood or glovebox (see image).

Steps required before opening the system:

  1. Stop particle production.
  2. Flush the system (see the corresponding section in the operating and maintenance guide).
  3. Stop the gas flow.
  4. Take out your sample following the procedure for your deposition system, taking the necessary precautions to prevent/reduce exposure to nanoparticles. Pay special attention when handling potentially pyrophoric nanoparticles.

System care

top

The system requires no regular maintenance by the end user, but cleaning the reactor periodically is recommended, especially after a change of electrode material. To prevent exposure to nanoparticles and to ensure safe and efficient cleaning, guidelines are set out in the operating and maintenance guide. The procedure to replace electrodes is also discussed there.

When to clean the reactor and replace the filter(s)

Clean the reactor whenever you replace the electrodes, the reactor is dirty, or after 100-1000 hours of operation (which in turn depends on your operating conditions: if making more particles, clean more often). If higher settings are needed to maintain a spark flow, it is possible that the reactor needs to be cleaned to prevent spark obstruction which can possibly lead to spark misfiring within the reactor system. If the pressure within the system increases even with the valves open, it is possible that the filter is full and needs to be replaced. Refer to the operating and maintenance guide for more information.

Safe handling and storage

The VSP-G1 is a scientific instrument, and must be handled with care. This section provides general guidelines on how to safely handle and store the system. Detailed methods are described in the operating and maintenance guide.

Alert Warning: Flush the system before opening the system.

Disassembly

The base unit with reactor installed has a combined weight of 22kg. For general handling, e.g. when assembling a setup, the reactor is conveniently be transported separately from the base unit. A mount plate / transport tool is provided for this purpose. See unmounting the reactor assembly of the operating and maintenance guide.

General disposal guidelines

Disposal of nanomaterials

Label and dispose cleaning materials contaminated with nanomaterials according to local rules and regulations. See also Filter replacement and disassembling and cleaning the reactor.

Retiring the VSP-G1

Local regulations and practices apply for disposal of electronic equipment. If the VSP-G1 has reached end of life, please contact VSParticle at support@vsparticle.com for recommendations on how to dispose of the system.

Transportation, packaging and storage

VSParticle recommends using the original packaging for transportation and shipping. Be sure to clean the reactor before shipping the reactor. The system should also be cleaned if it is not to be used for a longer period of time.

When shipping the system to VSParticle or one of it's representatives for any reason, return authorization (RA) is required. Contact your distributor. For systems supported by VSParticle, an RA form can be requested by sending an e-mail to support@vsparticle.com.

Troubleshooting

There are several possible issues that may arise during normal operation of the VSP-G1. The most common scenarios and errors are listed below.

Unstable operation

In some scenarios, unstable operation of the spark (wild fluctuations of the gap voltage, intermittent sparking and/or the system is slow to reach its set point) may be encountered. There are several causes for such instability to occur:

  • Glow mode is active. In this case, a weak arc is observed between both electrodes and the gap voltage drops suddenly. Please refer to the glow mode section below.
  • The carrier gas selected on the system differs from the carrier gas being used. Symptoms include wild fluctuations around the intended gap voltage set point and the motor in the system can be heard adjusting the gap all the time. Alternatively, the system may be very slow to reach the intended set point. To correct the issue, ensure the correct carrier gas is selected by following the instructions in the operating and maintenance guide.
  • Oxidation of the electrode material has occurred. This can result in fluctuations and/or intermittent sparking until the oxidised material has been ablated. To resolve this issue more quickly, increase the spark frequency by increasing the charging current set point and optionally the gap voltage (to avoid glow mode).

Glow mode

Occasionally the spark voltage drops suddenly, and a weak arc may be seen between the two electrodes. This is caused by a too high charging current. If this occurs directly after changing the spark voltage or spark current set point, the control system in the VSP-G1 will likely correct the situation. If not, the chosen set point cannot be maintained in spark mode. Increase the spark voltage or decrease the operating current to correct the situation.

Interlock errors

Message on display Cause Resolution
Interlock err. 7 Motor controller failure The motor controller reports a failure. This can be due to:
  • A blocked motor. Follow the instructions described in the VSP-G1 Operating and maintenance guide to dismount and disassemble the reactor. Then, verify that the electrode is free to move by manually rotating the gear located at the underside of the left end cap. Assemble and remount the reactor ensuring some space exists between both electrodes and power on the VSP-G1 to try again.
  • Motor controller communication interference or failure. Press both dials to clear the error and try again. Optionally, switch the machine off and on again. If the error persists, contact VSParticle.
  • Motor controller failure. Press both dials to clear the error and/or switch the device off and on again. If the error persists, contact VSParticle.
Interlock Error 8 Broken interlock switch One of the interlock switches is broken or has a broken connection. Verify both reactor mounting pins are properly secured and that the reactor is properly mounted on the base unit. Press both dials to clear the error and try again. If the error persists, contact VSParticle.
Interlock Error 9 Open interlock switch One or more interlock switches are in the open state. Verify both reactor mounting pins are properly secured and that the reactor is properly mounted on the base unit. Press both dials to clear the error and try again.

If you encounter any interlock error other than the ones listed above, please try to clear the interlock by pressing and releasing both dials and/or switch the VSP-G1 off and on again. If the error persists, please contact VSParticle via support@vsparticle.com.

Electrode alignment

When the VSP-G1 starts aligning electrodes during start-up, several sounds may be heard:

  • Quick hammering. This occurs occasionally when the electrode is in an unusual position (e.g. after maintenance/cleaning of the reactor has been performed) and cannot move in the expected direction. This is normal. The sounds should not last more than a second and the system corrects itself and completes the alignment procedure. If the sounds persist for more than a few seconds, switch off the VSP-G1 using the main power switch at the back. Follow the instructions for a blocked motor found under the Interlock err. 7 entry of the interlock errors section above.
  • Clicks. These occur at the boundaries of the adjustment range and are no cause for concern.
  • Vibrations or rattling. This can occur when there is tension on the gas inlet of the left end cap, restricting free movement or due to wear of the transmission. Check that the gas inlet is free to move. If these sounds occur regularly, contact VSParticle.


When the electrode alignment sequence fails, this is displayed as Interlock err. 7. Refer to the interlock error section above for possible resolutions.

When to contact VSParticle

  • If the system keeps reverting to interlock without apparent reason.
  • If sparks can be heard or seen in any location other than between the electrode tips.

Accessories and replacement parts

top

Use only VSParticle approved parts. All parts related to the reactor assembly must be examined and/or supplied by VSParticle. Contact us for more information.

Specifications of replaceable parts that can be obtained by other suppliers:
Fuses - The fuse holder accepts 5x20mm T2.5A/250V fuses.
Power cable - IEC-60320 compliant with C13 connector,
RS-232 cable - any standard (straight) RS-232 cable is suitable.

Appendices

top

VSP-G1 technical specifications

Additional information about the VSP-G1 operating window, electrical specifications and output can be found on the Technical Specifications page.