VSP-G1 Quick Start Guide
From VSPARTICLE - Docs
This is the Quick Start Guide for standard operation of a VSParticle Generator One (VSP-G1) unit. For first time operators, detailed information or troubleshooting, please refer to the User Manual.
A themed and printable version of the quick start guide is available for download here: File:VSP-G1 Quick Start Guide.pdf. A laminated copy of this version is included in the box in which the VSP-G1 is delivered.
Ensure a safe system
![]() |
Warning: Failure to verify the items below may result in compromised operation of the system, including exposure to nanoparticles, pressure build-up and performance issues. |
- Only use flexible tubing for the gas inlet when using the gas inlet labeled "Gas inlet only". This inlet will move during start up.
- Use conductive tubing for any gas outlet used as particle exhaust.
- Ensure the system is leak tight. A system that is not leak tight may expose the operator to nanoparticles during operation. Refer to the user manual section on leak tightness for more information.