This is the comprehensive User Manual for the VSParticle Generator One unit. The Quick Start Guide is available as a quick reference for standard operation of the G1 unit.
-
VSParticle Generator One unit (VSP-G1)
[INCLUDE ALSO CONTACT INFORMATION within a framed box]
Introduction
This user manual provides instructions for the set-up, installation, operation and maintenance of a VSParticle Generator One (VSP-G1) unit. Before operating the VSP-G1 unit, please read the safety section carefully and pay attention to the warnings provided throughout the manual, indicated by the following symbols:
(symbols used in the manual)
(Short index of manual and a 1 sentence summary of each chapter/section)
A quick start guide is also included for faster set-up as part of the standard operating procedure. For more information regarding safety and more complex operations as well as troubleshooting and maintenance, please refer to this Manual.
Technical Specifications
Power 110-240V AC Dimensions Casing ca. 52x30x20cm Reactor added height ca. 10cm Weight 19kg Gas inlet/outlet 10mm tubes (with Swagelok connectors) Display 16x2 characters Digital output RS232
Operating Window
Flow rate | 1-30 L/min |
Gas | Supported: Ar or N2 (recommended purity 5.0) Unsupported: He, Ne, Xe, Kr Contact VSParticle for the use of reactive gases such as air and H2. |
Electrode material | Comes with Cu electrodes. Various other metals (e.g. Ag, Au, Pt, W, Ni), semiconductors and carbon are possible. |
Primary particle size | 1 atom to 20 nm |
Ablation rate | ~0.01-100 mg/h (material dependent) |
Typical particle concentration | 108-1011 cm-3 |