VSP-G1 User Manual: Difference between revisions
From VSPARTICLE - Docs
Line 14: | Line 14: | ||
{| style="margin-left: auto; margin-right: auto; border: 1px solid orange;" | {| style="margin-left: auto; margin-right: auto; border: 1px solid orange;" | ||
|- | |- | ||
|'' | |''Power'' ||110-240V AC | ||
|- | |- | ||
|'' | |''Dimensions'' ||Casing ca. 52x30x20cm | ||
|- | |- | ||
| || | | ||Reactor added height ca. 10cm | ||
|- | |- | ||
| || | |''Weight'' ||19kg | ||
|- | |- | ||
|'' | |''Gas inlet/outlet'' ||10mm tubes (with Swagelok connectors) | ||
|- | |- | ||
| || | |''Display'' ||16x2 characters | ||
|- | |- | ||
|'' | |''Digital output'' ||RS232 | ||
|} | |} | ||
Revision as of 12:26, 7 September 2016
This is the comprehensive User Manual for the VSParticle Generator One unit. The Quick Start Guide is available as a quick reference for standard operation of the G1 unit.
-
VSParticle Generator One unit
[INCLUDE ALSO CONTACT INFORMATION within a framed box]
Introduction
Technical Specifications
Power | 110-240V AC |
Dimensions | Casing ca. 52x30x20cm |
Reactor added height ca. 10cm | |
Weight | 19kg |
Gas inlet/outlet | 10mm tubes (with Swagelok connectors) |
Display | 16x2 characters |
Digital output | RS232 |
Operating Window
Flow rate | 1-30 L/min |
Gas | Supported: Ar or N2 (recommended purity 5.0) |
Unsupported: He, Ne, Xe, Kr | |
Contact VSparticle for the use of reactive gases such as air and H2. | |
Electrode material | Comes with Cu electrodes. |
Various other metals (e.g. Ag, Au, Pt, W, Ni), semiconductors and carbon are possible. | |
Primary particle size | 1 atom to 20 nm |
Ablation rate | ~0.01-100 mg/h (material dependent) |
Typical particle concentration | 108-1011 cm-3 |