VSP-G1 Technical Specifications: Difference between revisions

From VSPARTICLE - Docs
(Created page with "== Unit Specifications == <blockquote style="padding: 1en;"> {| |''Power'' ||110-240V AC |- |''Dimensions'' ||Casing ca. 52x30x20cm |- | ||Reactor added height ca. 10cm |- |''...")
 
No edit summary
 
(9 intermediate revisions by 3 users not shown)
Line 1: Line 1:
== Unit Specifications ==
== Unit specifications ==
<blockquote style="padding: 1en;">
{| class="table table-bordered table-striped"
{|
|''Power'' ||110-240V AC
|-
|-
|''Dimensions'' ||Casing ca. 52x30x20cm
! Specification !! Value
|-
|-
| ||Reactor added height ca. 10cm
|''Mains supply voltage'' ||110-240V AC
|-
|-
|''Weight'' ||19kg
|''Mains frequency'' ||50-60Hz
|-
|''Power'' ||175 V-A
|-
|''Dimensions'' ||Casing ca. 52x30x20cm <br />Reactor added height ca. 10cm
|-
|''Weight'' ||Base unit: 16kg<br />
Reactor assembly: 6kg<br />
Mount plate: 3kg
|-
|-
|''Gas inlet/outlet'' ||10mm tubes (with Swagelok connectors)
|''Gas inlet/outlet'' ||10mm tubes (with Swagelok connectors)
Line 16: Line 22:
|''Digital output'' ||RS232
|''Digital output'' ||RS232
|}
|}
</blockquote>


== Operating Window ==
== Operating window ==
<blockquote style="padding: 1en;">
{| class="table table-striped table-bordered"
{|
! Parameter !! Value
|-valign="top"
|-
|''Flow rate'' ||1-30 L/min
|''Flow rate'' ||1-30 L/min
|-valign="top"
|-
|''Gas'' ||Supported: Ar or N<sub>2</sub> (recommended purity 5.0) <br> Unsupported: He, Ne, Xe, Kr <br> Contact VSParticle for the use of reactive gases such as air and H<sub>2</sub>.
|''Gas'' ||Supported: Ar or N<sub>2</sub> (recommended purity 5.0) <br> Unsupported: He, Ne, Xe, Kr <br> {{Notification|Contact VSParticle| for the use of reactive gases such as air and H<sub>2</sub>.}}
|-valign="top"
|-
|''Electrode material'' ||Comes with Cu electrodes. <br> Various other metals (e.g. Ag, Au, Pt, W, Ni), semiconductors and carbon are possible.
|''Electrode material'' ||Comes with Cu electrodes. <br> Various other metals (e.g. Ag, Au, Pt, W, Ni), semiconductors and carbon are possible.
|-valign="top"
|-
|''Primary particle size'' ||1 atom to 20 nm
|''Primary particle size'' ||1 atom to 20 nm
|-valign="top"
|-
|''Ablation rate'' ||~0.01-100 mg/h (material dependent)
|''Ablation rate'' ||~0.01-100 mg/h (material dependent)
|-valign="top"
|-
|''Typical particle concentration'' ||10<sup>8</sup>-10<sup>11</sup> cm<sup>-3</sup>
|''Typical particle concentration'' ||10<sup>8</sup>-10<sup>11</sup> cm<sup>-3</sup>
|}
|}
</blockquote>
 
== External circuit terminals ==
{| class="table table-bordered table-striped"
!Terminal
|RS-232
|-
!Description
|Serial connector
|-
!Connector
|F-DE9 (sub-D, 9 pin)
|-
!Rating/specification
|RS-232, only pins 2, 3, 5 connected
|-
!Voltage rating
|Isolated 2.5kV
|}
 
== Environmental Conditions ==
The VSP-G1 is designed to be safe under the following normal environmental conditions:
* Indoor use;
* Altitude up to 2,000 m;
* Temperature 5 &deg;C to 40 &deg;C;
* Maximum relative humidity 80 % for temperatures up to 31 &deg;C decreasing linearly to 50 % relative humidity at 40 &deg;C;
* Mains supply voltage fluctuations up to &plusmn;10 % of the nominal voltage;
* Transient overvoltages up to the levels of overvoltage category II;
* Temporary overvoltages occurring on the mains supply; and
* Pollution degree 2.

Latest revision as of 12:21, 5 April 2017

Unit specifications

Specification Value
Mains supply voltage 110-240V AC
Mains frequency 50-60Hz
Power 175 V-A
Dimensions Casing ca. 52x30x20cm
Reactor added height ca. 10cm
Weight Base unit: 16kg

Reactor assembly: 6kg
Mount plate: 3kg

Gas inlet/outlet 10mm tubes (with Swagelok connectors)
Display 16x2 characters
Digital output RS232

Operating window

Parameter Value
Flow rate 1-30 L/min
Gas Supported: Ar or N2 (recommended purity 5.0)
Unsupported: He, Ne, Xe, Kr

Notification: Contact VSParticle for the use of reactive gases such as air and H2.

Electrode material Comes with Cu electrodes.
Various other metals (e.g. Ag, Au, Pt, W, Ni), semiconductors and carbon are possible.
Primary particle size 1 atom to 20 nm
Ablation rate ~0.01-100 mg/h (material dependent)
Typical particle concentration 108-1011 cm-3

External circuit terminals

Terminal RS-232
Description Serial connector
Connector F-DE9 (sub-D, 9 pin)
Rating/specification RS-232, only pins 2, 3, 5 connected
Voltage rating Isolated 2.5kV

Environmental Conditions

The VSP-G1 is designed to be safe under the following normal environmental conditions:

  • Indoor use;
  • Altitude up to 2,000 m;
  • Temperature 5 °C to 40 °C;
  • Maximum relative humidity 80 % for temperatures up to 31 °C decreasing linearly to 50 % relative humidity at 40 °C;
  • Mains supply voltage fluctuations up to ±10 % of the nominal voltage;
  • Transient overvoltages up to the levels of overvoltage category II;
  • Temporary overvoltages occurring on the mains supply; and
  • Pollution degree 2.